IV. S E T U P A N D O P E R A T I N G I N S T R U C T I O N S - C O N T .
27
NOTE: If it is desired to deactivate the oxygen sensor, select it in the Alarm 2 menu. Scroll the
lower limit down to off. The function is now inoperative. The function may be reactivated at any time
by pressing it and scrolling the low limit up. It is now reactivated. This allows the ventilator to be
used without the oxygen sensor. When this sensor is off, it may be disconnected while the ventilator is
in use, without causing an alarm.
3- SUMMARY OF OPERATING INSTRUCTIONS
Turn on.
Select mode.
Set primary functions in MAIN Menu.
Press AUTO SET.
Verify proper operation.
Connect to patient.
Make patient appropriate adjustments to settings.
Press the alarm menus key and adjust alarm limits if desired.
IMPORTANT NOTES:
1- Flashing I, E or I/E Ratio display indicates a timing limit violation and can be permanently silenced only by
adjusting the settings to bring them into permissible ranges.
2- In Constant Flow mode with CPAP, Backup is inoperative.