10
3.
Continue to assist in the evacuation of air from garment, working from top to bottom.
4.
Once the garment feels loose enough, you can unzip the garment all the way to bottom and remove.
NOTE: Repeat above steps if bilateral garments are used!
ADJUSTING GRADIENT PRESSURES
1.
Model # SC-3008 Sequential Circulators provide total Calibrated Gradient Pressure.
2.
All gradients have been carefully calibrated at the factory resulting in the highest pressure in chambers
#1 & #2 graduated down to the lowest in chambers #7 & #8.
3.
The main PRESSURE ADJUSTMENT KNOB (#2) sets the pressure for sections #1 & #2 of the sleeve. As it
is adjusted, it will also raise or lower in proportion, the remaining sections.
4.
In the unlikely event you would wish to change the gradient settings calibrated at the factory, it is pos-
sible to do so by using the “over-ride” adjustments (#8) on the underside of the pump. (See
“Overriding Pre-set Gradient Pressures” next page)
5.
We described earlier how the Pressure Gauge responds to filling pressures with a brief “dip” before
raising back up to the actual pressure.
NOTE: Repeat above steps if bilateral garments are used!