12 

 

Garment Disinfection 

 

  Disconnect from pump and open to expose all sides 

  Prepare a solution of 1/3 cup of laundry detergent per 1 gallon of warm 

water (20 mL laundry detergent per 1 L water) in a sink or container 
large enough to hold the garment 

  Place garment in solution but do not submerge or place connectors in 

water as this will damage the device 

  Soak for 30 minutes with mild agitation every 5-10 minutes - hard to 

remove soil may require hand washing using a soft clean cloth while 
garment is in solution 

  Rinse with warm water and allow to air dry 

  Repeat previous steps using a solution of 1 cup of bleach per 1 gallon 

of warm water (60 mL bleach per 1 L warm water) 

 

Storing and Transporting 

 

  Keep and reuse packaging for transporting the device 

  Store in a dry location away from a source of heat and free of pests 

 

Servicing and Repairs 

 

  Contact Bio Compression Systems for servicing 

– there are no user 

serviceable parts 

  Tampering, modifying, or dismantling this device in any way voids the 

warranty 

  When contacting Bio Compression Systems, please have your model 

number and serial number ready 

 

Troubleshooting 

 

Pump does not turn on: 

1.  Check to see if the pump is plugged in 
2.  Unplug and examine power cord for damaged - if damaged, contact 

Bio Compression Systems 

3.  Check circuit breaker to make sure outlet has power 

4. 

Contact Bio Compression Systems

 

 

Garment does not deflate: 

1.  Check garment connection to pump 
2.  Check garment hose for damage, kinks, or twists 
3.  Check garment for damage 
4.  Contact Bio Compression Systems 

 

Summary of Contents for SC-2004-DL

Page 1: ...SEQUENTIAL CIRCULATOR Operating Instructions Quality Medical Products Since 1983...

Page 2: ...scription and Operating Principle 4 Warnings and Precautions 7 Operating Instructions 8 Accessories 13 Product Specifications 15 Environmental Specifications 16 Symbol Glossary 18 Information for Dist...

Page 3: ...the limb including cellulitis without appropriate antibiotic coverage The presence of lymphangiosarcoma Suspicion or confirmation of the presence of Deep Vein Thrombosis DVT Inflammatory phlebitis or...

Page 4: ...er study from the physician for their records Device Description and Operating Principle Sequential Circulators provide gradient pneumatic compression for the treatment of lymphedema and associated ve...

Page 5: ...attention Focus therapy doubles the inflation time of two adjacent chambers SC 2004 DL or three adjacent chambers SC 2008 DL during the first ten minutes of treatment Play Pause The Play Pause button...

Page 6: ...tors 2 LED Display 3 UP Button 4 DOWN Button 5 SET Button 6 Power On Off Button 7 PLAY PAUSE Button 8 Garment Connector Bar Port 9 Auxiliary Connector Bar Port shown with Blocker Bar The SC 2008 DL ha...

Page 7: ...used no closer than 12 30 cm to any part of the device including the power cord otherwise degradation of the performance of this equipment could result Use of this equipment adjacent to or stacked wi...

Page 8: ...pump settings unless directed by a physician Do not carry or suspend the device using tubing valves or the power cord as handles Do not submerge the device or allow liquids to enter the device Never...

Page 9: ...OWN button to adjust pressure and press SET button when finished SEt will appear in LED Display and Cycle Time LED indicator will be lit Use UP or DOWN button to adjust cycle time and press SET button...

Page 10: ...ent see below At the end of a timed treatment your pump will turn off To turn off the pump during a treatment press the Power On Off button HOld will show in the LED and then the pump will turn off Sq...

Page 11: ...owed by the hours of use since the last reset Information will be displayed until Play Pause is hit again To Check Pressure Setting While Pump is Running Hold the Down button until the pressure settin...

Page 12: ...L warm water Storing and Transporting Keep and reuse packaging for transporting the device Store in a dry location away from a source of heat and free of pests Servicing and Repairs Contact Bio Compre...

Page 13: ...m Shoulder Sleeve Large Right GV 3000 4 Chamber Vest GS 3045 H 4 Chamber Leg Sleeve Half GS 3045 S 4 Chamber Leg Sleeve Small GS 3045 M 4 Chamber Leg Sleeve Medium GS 3045 L 4 Chamber Leg Sleeve Large...

Page 14: ...ve A8 3045 S 8 Chamber Adjustable Leg Sleeve Small A8 3045 M 8 Chamber Adjustable Leg Sleeve Medium A8 3045 L 8 Chamber Adjustable Leg Sleeve Large GWA8 3045 S 8 Chamber Wide Adjustable Leg Sleeve Sma...

Page 15: ...le Time 60 120 seconds in 15 second increments Treatment Time Continuous or 10 120 minutes in 5 minute increments Pressure Range 10 120 mmHg Precision 1 mmHg Accuracy 20 Features Adjustable Cycle Time...

Page 16: ...he pump when not in use unplugging electronic devices when they are not being used saves electricity Do not clean garment soiled this minimizes the consumables use Reuse packaging for storing and tran...

Page 17: ...pply network that supplies building used for domestic purposes Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Electromagnetic Immun...

Page 18: ...tric shock Restricted to sale by or on the order of a physician Complies with the Waste Electrical and Electronic Equipment Directive WEEE Directive Serial number Complies the European Medical Device...

Page 19: ...reset the treatment compliance meter and return the pump to factory settings Begin with the device turned off and plugged in Press and hold the UP DOWN and SET buttons simultaneously for 5 seconds LE...

Page 20: ...20 L 284 C EN 2021 09...

Reviews: