To select calf mode 

 
• Begin with device turned off 
• Press and hold DOWN button and momentarily press power button 
• Elapsed time will appear followed by “CAF” 
• Release DOWN button 
• Pressure will appear (50 mmHg) and device will start running 
 

To select foot mode 

 
• Begin with device turned off 
• Press and hold UP button and momentarily press power button 
• Elapsed time will appear followed by “FOO” 
• Release UP button 
• Pressure will appear (120 mmHg) and device will start running 
 

To change pressure 

 
• When display shows “0” during cycle, press and hold UP button until pressure appears 
• Adjust using UP or DOWN button 
• Press “Power On/Off” button to begin cycle 
 

Cleaning 

 
The pump, garment, and tubing can be wiped down using a damp (not wet) soft cloth 
while unplugged 

– if more thorough pump cleaning or garment disinfection is desired, 

use the following directions. 
 

Pump, tubing, and calf garment cleaning 

 
• Unplug and wipe down using a damp (not wet) soft cloth with mild antibacterial soap 
as needed 
• Do not use bleach 
 

Foot and cryo garment disinfection 

 
• Disconnect from pump and open to expose all sides 
• Prepare a solution of 1/3 cup of laundry detergent per 1 gallon of warm water (20 mL 
laundry detergent per 1 L water) in a sink or container large enough to hold the garment 
• Place garment in solution but do not submerge or place connectors in water as this will 
damage the device 
• Soak for 30 minutes with mild agitation every 5-10 minutes - hard to remove soil may 
require hand washing using a soft clean cloth while garment is in solution 
• Rinse with warm water and allow to air dry 
• Repeat previous steps using a solution of 1 cup of bleach per 1 gallon of warm water 
(60 mL bleach per 1 L warm water) 

Summary of Contents for IC-1545-DL Multi-Flo DVT Combo

Page 1: ...Intermittent Pneumatic Compression Device Operating Instructions Quality Medical Products Since 1983...

Page 2: ...2...

Page 3: ...n 4 Intended Use 4 Device Description and Operating Principle 5 Warnings and Precautions 6 Operating Instructions 8 Accessories 11 Product Specifications 11 Environmental Specifications 12 Symbol Glos...

Page 4: ...Cold Compression Intended to help reduce pain and swelling after surgery Enhances faster healing and tissue regeneration required by the body Contraindications Use of this device is contraindicated fo...

Page 5: ...atment A physician is required to prescribe these settings but general guidelines are listed below It is ultimately the physician s responsibility to prescribe the setting and it should be written on...

Page 6: ...of electric shock burns fire injury or improper treatment read the entire instruction manual before operating this device Use of accessories or a power cord not specified or provided by Bio Compressi...

Page 7: ...ing heat or cold When Using This Product Examine the device accessories and power cord for damage before using or cleaning Handle garments with care do not fold or crease use near a heat source handle...

Page 8: ...y surface pump must be close enough for the controls to be reached during use Plug power cord into outlet Remove garments from plastic bag unroll and spread flat Place gel insert flat in freezer for 4...

Page 9: ...lugged if more thorough pump cleaning or garment disinfection is desired use the following directions Pump tubing and calf garment cleaning Unplug and wipe down using a damp not wet soft cloth with mi...

Page 10: ...Check to see if the pump is plugged in 2 Unplug and examine power cord for damaged if damaged contact Bio Compression Systems 3 Check circuit breaker to make sure outlet has power 4 Contact Bio Compr...

Page 11: ...tion IP20 Mains Isolation Unplug Mode of Operation Continuous Essential Performance The pump s cyclical inflation and deflation of the garment s Cycle Time 60 10 seconds Inflation 15 seconds calf mode...

Page 12: ...ices when they are not being used saves electricity Unplug the pump when not in use unplugging electronic devices when they are not being used saves electricity Do not clean garment soiled this minimi...

Page 13: ...nd those directly connected to the public low voltage power supply network that supplies building used for domestic purposes Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicke...

Page 14: ...c shock Restricted to sale by or on the order of a physician Complies with the Waste Electrical and Electronic Equipment Directive WEEE Directive Serial number Complies the European Medical Device Reg...

Page 15: ...7074 USA Phone 1 201 939 0716 Toll Free Phone US 800 888 0908 E mail biosystems biocompression com Website www biocompression com When contacting us please have your model number and serial number rea...

Page 16: ...16 L 008 I EN 2020 04...

Reviews: