9
To select calf mode
• Begin with device turned off
• Press and hold DOWN button and momentarily press power button
• Elapsed time will appear followed by “CAF”
• Release DOWN button
• Pressure will appear (50 mmHg) and device will start running
To select foot mode
• Begin with device turned off
• Press and hold UP button and momentarily press power button
• Elapsed time will appear followed by “FOO”
• Release UP button
• Pressure will appear (120 mmHg) and device will start running
To change pressure
• When display shows “0” during cycle, press and hold UP button until pressure appears
• Adjust using UP or DOWN button
• Press “Power On/Off” button to begin cycle
Cleaning
The pump, garment, and tubing can be wiped down using a damp (not wet) soft cloth
while unplugged
– if more thorough pump cleaning or garment disinfection is desired,
use the following directions.
Pump, tubing, and calf garment cleaning
• Unplug and wipe down using a damp (not wet) soft cloth with mild antibacterial soap
as needed
• Do not use bleach
Foot and cryo garment disinfection
• Disconnect from pump and open to expose all sides
• Prepare a solution of 1/3 cup of laundry detergent per 1 gallon of warm water (20 mL
laundry detergent per 1 L water) in a sink or container large enough to hold the garment
• Place garment in solution but do not submerge or place connectors in water as this will
damage the device
• Soak for 30 minutes with mild agitation every 5-10 minutes - hard to remove soil may
require hand washing using a soft clean cloth while garment is in solution
• Rinse with warm water and allow to air dry
• Repeat previous steps using a solution of 1 cup of bleach per 1 gallon of warm water
(60 mL bleach per 1 L warm water)