In the event the pump stops working

 (e.g., power failure), release pressure by 

disconnecting the garment.

 

 

Any serious incident that has occurred in relation to the device must be reported 
to Bio Compression Systems. 

In the European Union (EU), incidents should also be 

reported to the competent authority of the Member State in which the user and / or 
patient is established.

 

 

Keep out of reach of children and pets. 

 

Operating Instructions 

 
The patient is the intended user and can safely use all functions. 
 

Preparing the Device for Use 

 
• Remove garments and cardboard underneath from box 
• Lift pump out of box and remove protective end caps – save packaging for transport 
and storage 
• Place pump on a flat sturdy surface – pump must be close enough for the controls to 
be reached during use 
• Plug power cord into outlet 
• Remove garments from plastic bag, unroll, and spread flat 
• Place gel insert flat in freezer for 4 hours - do not place directly on metal rack as this 
could damage inset  
 

Connecting the Garments 

 
• Connect tubing to garment valve by turning clockwise until secure 
• Attach Quick-Connect connector at end of tubing to the pump by firmly pushing 
inwards until it snaps in 
 

Putting the Garments On 

 
• Apply garments and secure using Velcro fasteners 
 

Operating the Device 

 
• Press “Power On/Off” button to turn on 
• The device is preset for calf mode at a pressure of 50 mmHg (to change, see below) 
• If garment does not properly inflate, alarm will sound – check connection and tubing 
for obstructions then turn pump off and on 
• Upon the completion of treatment or to stop treatment, press “Power On/Off” button to 
turn off 
 

 

 

Summary of Contents for IC-1545-DL Multi-Flo DVT Combo

Page 1: ...Intermittent Pneumatic Compression Device Operating Instructions Quality Medical Products Since 1983...

Page 2: ...2...

Page 3: ...n 4 Intended Use 4 Device Description and Operating Principle 5 Warnings and Precautions 6 Operating Instructions 8 Accessories 11 Product Specifications 11 Environmental Specifications 12 Symbol Glos...

Page 4: ...Cold Compression Intended to help reduce pain and swelling after surgery Enhances faster healing and tissue regeneration required by the body Contraindications Use of this device is contraindicated fo...

Page 5: ...atment A physician is required to prescribe these settings but general guidelines are listed below It is ultimately the physician s responsibility to prescribe the setting and it should be written on...

Page 6: ...of electric shock burns fire injury or improper treatment read the entire instruction manual before operating this device Use of accessories or a power cord not specified or provided by Bio Compressi...

Page 7: ...ing heat or cold When Using This Product Examine the device accessories and power cord for damage before using or cleaning Handle garments with care do not fold or crease use near a heat source handle...

Page 8: ...y surface pump must be close enough for the controls to be reached during use Plug power cord into outlet Remove garments from plastic bag unroll and spread flat Place gel insert flat in freezer for 4...

Page 9: ...lugged if more thorough pump cleaning or garment disinfection is desired use the following directions Pump tubing and calf garment cleaning Unplug and wipe down using a damp not wet soft cloth with mi...

Page 10: ...Check to see if the pump is plugged in 2 Unplug and examine power cord for damaged if damaged contact Bio Compression Systems 3 Check circuit breaker to make sure outlet has power 4 Contact Bio Compr...

Page 11: ...tion IP20 Mains Isolation Unplug Mode of Operation Continuous Essential Performance The pump s cyclical inflation and deflation of the garment s Cycle Time 60 10 seconds Inflation 15 seconds calf mode...

Page 12: ...ices when they are not being used saves electricity Unplug the pump when not in use unplugging electronic devices when they are not being used saves electricity Do not clean garment soiled this minimi...

Page 13: ...nd those directly connected to the public low voltage power supply network that supplies building used for domestic purposes Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicke...

Page 14: ...c shock Restricted to sale by or on the order of a physician Complies with the Waste Electrical and Electronic Equipment Directive WEEE Directive Serial number Complies the European Medical Device Reg...

Page 15: ...7074 USA Phone 1 201 939 0716 Toll Free Phone US 800 888 0908 E mail biosystems biocompression com Website www biocompression com When contacting us please have your model number and serial number rea...

Page 16: ...16 L 008 I EN 2020 04...

Reviews: