8
In the event the pump stops working
(e.g., power failure), release pressure by
disconnecting the garment.
Any serious incident that has occurred in relation to the device must be reported
to Bio Compression Systems.
In the European Union (EU), incidents should also be
reported to the competent authority of the Member State in which the user and / or
patient is established.
Keep out of reach of children and pets.
Operating Instructions
The patient is the intended user and can safely use all functions.
Preparing the Device for Use
• Remove garments and cardboard underneath from box
• Lift pump out of box and remove protective end caps – save packaging for transport
and storage
• Place pump on a flat sturdy surface – pump must be close enough for the controls to
be reached during use
• Plug power cord into outlet
• Remove garments from plastic bag, unroll, and spread flat
• Place gel insert flat in freezer for 4 hours - do not place directly on metal rack as this
could damage inset
Connecting the Garments
• Connect tubing to garment valve by turning clockwise until secure
• Attach Quick-Connect connector at end of tubing to the pump by firmly pushing
inwards until it snaps in
Putting the Garments On
• Apply garments and secure using Velcro fasteners
Operating the Device
• Press “Power On/Off” button to turn on
• The device is preset for calf mode at a pressure of 50 mmHg (to change, see below)
• If garment does not properly inflate, alarm will sound – check connection and tubing
for obstructions then turn pump off and on
• Upon the completion of treatment or to stop treatment, press “Power On/Off” button to
turn off