KBF / KBF P (E5.2) 02/2012
page 74/103
When using the communication software APT-COM™ 3 DataControlSystem (option, chap.
15.1) via the Ethernet interface of the climatic chamber for data acquisition, the alarm is not
automatically transmitted to the APT-COM™ protocol.
Set the tolerance limits for recording limit value excesses separately in APT-COM™ 3.
15.6 Water protected internal socket (option – not for UL units)
The internal socket is splash proof.
IP system of protection 65 230 V 1N ~ 50-60 Hz
Charge max. 500 W
Maximum permitted operating temperature with this option: 50 °C /
122 °F.
WARNING
Exceeding the permitted maximum temperature.
Electrical hazard.
Danger of death.
Damage to the internal socket.
∅
Do NOT exceed the temperature set-point of 50 °C /
122 °F.
Set the safety controller to 50 °C /
122 °F.
With optional temperature safety device, class 3.3, set the mechanical thermostat class
3.1 to 50 °C /
122 °F.
Heat emission of electrical devices connected inside the chamber may modify the temperature
and humidity range.
CAUTION
Risk of short circuit.
Damage to the unit.
Use the supplied plug only (IP protection type 66). Plug it in and turn it to secure con-
tact.
If the socket is not used, close the lift-up lid and turn it to secure.
Summary of Contents for KBF 115 (E5.2)
Page 5: ...KBF KBF P E5 2 02 2012 page 5 103 Product registration ...
Page 98: ...KBF KBF P E5 2 02 2012 page 98 103 19 7 Dimensions KBF 115 ...
Page 99: ...KBF KBF P E5 2 02 2012 page 99 103 19 8 Dimensions KBF 240 KBF P 240 ...
Page 100: ...KBF KBF P E5 2 02 2012 page 100 103 19 9 Dimensions KBF 720 KBF P 720 ...