12. Humidity system
The humidity system is turned on with the humidity switch (4) located in the right lateral control panel.
The climatic chamber KBF / KBF P is equipped with a capacitive humidity sensor. This results in a control
accuracy of up to
±
3 % r.H. of the set point. The temperature-humidity diagrams (Figure 21) show the
possible working ranges for humidity.
The preset temperature and humidity values should be situated within the optimum range
(hatched range in Figure 21). Only within this area will the unit not be exposed to excessive
moisture due to condensation.
In the short-term set points outside the optimum range can also be targeted. The control accu-
racies of
±
3 % r.H., however, cannot be guaranteed in this case.
0
10
20
30
40
50
60
70
80
90
100
10
20
30
40
50
60
70
80
90
100
% r.H.
°C
A
B
0
10
20
30
40
50
60
70
80
90
100
10
20
30
40
50
60
70
80
90
100
% r.H.
°C
A
B
C
KBF
KBF P without light cassettes
0
10
20
30
40
50
60
70
80
90
100
10
20
30
40
50
60
70
80
90
100
% r.H.
°C
A
B
C
0
10
20
30
40
50
60
70
80
90
100
10
20
30
40
50
60
70
80
90
100
% r.H.
°C
A
C
KBF P 240 with illumination
KBF P 720 with illumination
Figure 21: Temperature-Humidity diagrams
Summary of Contents for KBF 115 (E5.2)
Page 5: ...KBF KBF P E5 2 02 2012 page 5 103 Product registration ...
Page 98: ...KBF KBF P E5 2 02 2012 page 98 103 19 7 Dimensions KBF 115 ...
Page 99: ...KBF KBF P E5 2 02 2012 page 99 103 19 8 Dimensions KBF 240 KBF P 240 ...
Page 100: ...KBF KBF P E5 2 02 2012 page 100 103 19 9 Dimensions KBF 720 KBF P 720 ...