Then the settings for the program course are defined. Two parameters can be set:
•
Program delay time, i.e. a defined time before a program starts. It can be entered with a precision of
1 minute, and its maximum value is 99.59 (99 hs 59 min). If the value is 00.00, the program will start
immediately. During the program delay time, all chamber functions (heating, refrigeration, and fan) are
off.
•
Number of program cycles, i.e. the desired number of program repeats. Values from 1 to 99 can be
entered. If the program is not going to be repeated, enter the value “0”. For infinite repeats enter the
value “–1”. The program is repeated as a whole, it is not possible to repeat individual sections.
In the last step the selected program is started. These steps must be carried out subsequently.
Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before
starting a program.
Step 1 – Program selection (only with program type “2 programs” set):
Normal Display
Press program key
Display 1 shows
e.g. 1
(actual selection of the program)
Display 2 shows
SEL.PRG.
(select program 1 or 2)
Enter program number 1 or 2 using arrow
keys
Value is shown in display 1.
Next step – entry of program course settings
Press program key
Display 1 shows
e.g. 00.00
(entered delay time hh.mm)
Display 2 shows
RUN TIME
(enter delay time of program start)
Set delay time in hh.mm using arrow keys
Value is shown in display 1.
Press program key
Display 1 shows
e.g. -1
(actual selection of the number of program cycles)
Display 2 shows
REPEAT
(enter number of program cycles)
Select number of cycles -1, 0, 1 etc. using
arrow keys
Value is shown in display 1.
Last step – program start:
Press program key
Display 1 shows
e.g. 1
(selected program)
Display 2 shows
RUN PRG.
(Question: start selected program?)
Press program key
Display 1 shows
e.g. 25.5
C
actual temperature value
Display 2 shows
P01:S01 01:49:39
(time running backwards)
(actual program P01, actual section S01, and remaining
time of program section S01)
Program is running. The green LED (3d) lights up.
Summary of Contents for KB 23
Page 75: ...KB 23 KB 23 UL E3 1 02 2019 page 75 87 20 7 KB KB UL 23 dimensions...
Page 77: ...KB 23 KB 23 UL E3 1 02 2019 page 77 87...
Page 78: ...KB 23 KB 23 UL E3 1 02 2019 page 78 87...
Page 80: ...KB 23 KB 23 UL E3 1 02 2019 page 80 87...
Page 81: ...KB 23 KB 23 UL E3 1 02 2019 page 81 87 22 Product registration...