KBF-S / KBF-S-UL (E6) 11/2021
page 28/95
The freshwater can outlet must NOT be placed on or above the rear panel of the chamber.
DANGER
Electrical hazard by water entering the chamber via the rear panel.
Deadly electric shock. Damage to the chamber.
Attach the regular support ONLY on the left side of the chamber.
Attach the magnetic support adjustable in height (option) ONLY on the front or on the
left side of the chamber.
∅
Do NOT place the freshwater can tap on or above the rear panel of the chamber
To operate the chamber with humidity, the freshwater can tap must be open
(chap. 6).
Do not firmly close the lid of the freshwater can to allow air to enter the freshwater
can. As a result, obstruction of the water supply due to negative pressure inside the
can is avoided.
Figure 12: Freshwater can with open tap
4.2 Condensate collection pan
The emerging condensate is collected in a condensate collection pan which is mounted under the chamber.
If necessary, the pan can be removed and emptied at any time.
KBF-S / KBF-S-UL 115
KBF-S / KBF-S-UL 240, 720, 1020
Figure 13: Condensate collection pan