KBF-S / KBF-S-UL (E6) 11/2021
page 34/95
Menu
Required access level
Functions
Service
“Service”
•
Configuration settings (only for BINDER Service)
•
Password changing for User and Admin
USB
Export: Any user
Import: “Admin”
•
Export of configuration, logger, and service data
•
Import of configuration data
Unless noted otherwise, the figure in this manual show the functional range, which is available for the user
with “Admin” authorization level
Note: When specifying the path to the respective function, the possibly required entry of a password is not
listed
5.2 Performance during and after power failures
During a power failure, all controller functions are shut down.
After the power returns, all functions return to the same status the chamber had before power failure. The
set-points are immediately resumed.
If during power failure an alarm has occurred (tolerance range, safety controller etc.), confirm the alarm.
See chap. 13.
6. Start up
After connecting the power supply and filling the freshwater can, turn on the chamber by its main power
switch (1). The lit pilot lamp shows the chamber is ready for operation.
The controller shows normal display and controls temperature and humidity to the last entered values.
For operation with humidity, the humidifying and dehumidifying system must be activated (chap. 8).
For operation with humidity:
•
Activate the humidifying and dehumidifying system (chap. 8).
•
Open the freshwater can tap (chap. 4.1.3)
•
Do not firmly close the lid of the freshwater can to allow air to enter the freshwater can. As a result,
obstruction of the water supply due to negative pressure inside the can is avoided.
After the first turning on of the chamber or after an interruption of the power supply the relative humidity will
increase after a delay of about 20 minutes. During this period, the relative humidity can drop considerably.
Warming chambers may release odors in the first few days after commissioning. This is not a quality defect.
To reduce odors quickly we recommend heating up the chamber to its nominal temperature for one day
and in a well-ventilated location.
WARNING: If customer should use a BINDER chamber running in non-supervised continu-
ous operation, we strongly recommend in case of inclusion of irrecoverable specimen or
samples to split such specimen or samples and store them in at least two chambers, if this is
feasible.
Summary of Contents for 9020-0366
Page 81: ...KBF S KBF S UL E6 11 2021 page 81 95 25 7 Dimensions Dimensions size 115...
Page 82: ...KBF S KBF S UL E6 11 2021 page 82 95 Dimensions size 240...
Page 83: ...KBF S KBF S UL E6 11 2021 page 83 95 Dimensions size 720...
Page 84: ...KBF S KBF S UL E6 11 2021 page 84 95 Dimensions size 1020...
Page 86: ...KBF S KBF S UL E6 11 2021 page 86 95...
Page 87: ...KBF S KBF S UL E6 11 2021 page 87 95...