14.6 Zero-voltage relay outputs via operation lines (option)
With this option, operation lines 2 to 5 serve to switch any device connected to the zero-voltage relay
output (DIN sockets (8) and (9) located in the lateral control panel). They can be programmed ON/OFF in
Manual Mode (chap. 8) as well as in Program Mode (AUTO, chap. 9) via operation lines 2 to 5.
Connection for operation lines 2 and 3 occurs via DIN socket (8), connection for operation lines 4 and 5
via DIN socket (9) in the lateral control panel:
OUTPUT TRACK 2+3
24V/MAX.2,5A
OUTPUT TRACK 4+5
24V/MAX.2,5A
Figure 12: Pin configuration of DIN sockets (8) left and (9) right
DIN socket (8):
Operation line 2
Operation line 3
Pin 1: Pin
Pin 2: Make
Pin 4: Pin
Pin 5: Make
DIN socket (9):
Operation line 4
Operation line 5
Pin 1: Pin
Pin 2: Make
Pin 4: Pin
Pin 5: Make
Maximum loading capacity of the switching contacts: 24V AC/DC – 2.5 A
DANGER
Electrical hazard.
Danger of death.
Damage to switching contacts and connection socket.
∅
Do NOT exceed the maximum switching load of 24V AC/DC – 2.5A.
∅
Do NOT connect any devices with a higher loading capacity.