MK (E3) 07/2009
page 31/70
Hit the PGM button. An inquiry screen appears offering to enter or delete individual program sections:
ZP-Abschnitt Abs. Nr. 5
insert
delete
In this view, new program lines can be entered or deleted:
new
New lines are added below in the table
insert
New lines are added above a previously selected line
delete
Individual lines that have been selected previously are deleted
Create as many lines, i.e. program sections, as desired. In a next step, values can be entered into these
lines. It is as well possible to add supplementary lines later on or to delete individual lines at any moment.
PGM
08:43:55 15.02.02
PGM
08:43:55 15.12.07
To enter values, select the corresponding line via the
arrow keys.
Hit the ENTER button. The
program editor
appears.
Enter the individual values of the selected program section.
Program editor Abs.Nr. 6
****.*
Setpoint 1
+100.0
FAN
Operating contacts
Time
00:45:00
Repeat Section
5
Repeat Number
10
Tol.-band min.
-1999.0
Tol.-band max.
+9999.0
Parameter set
1
-- Temperature at the start of the program section
-- Fan speed in % (no function)
-- Operating
contacts
(operation
lines)
(1
=
Bedew protection)
-- Duration of the program section
-- No. of start section in case of repeat cycles
-- No. of repeats in case of repeat cycles
-- Temperature limits (maximum / minimum temperature)
In case of exceeding: temporary program stop.
-- Pre-selected value (Do NOT change!)
Select the parameters via the arrow keys and confirm by ENTER.
Then enter the values via the arrow keys, and confirm the entry by ENTER.
For a negative set-point entry, enter the numerical value first, and only then the negative sign
(-).
With set-point type “
Limit
”, adapt the safety controller (chap. 11.2) to the highest temperature
set-point value of the program actually used. Check the safety controller for each temperature
program and adapt it if necessary. Set the safety controller set-point by about 5°C to 10°C
above the highest temperature set-point of the program.
Recommended setting: set-point type “
Offset
” with safety controller set-point 5°C.
In case of the optional over-/under temperature safety device (chap. 11.3), check also the
temperature limits entered there, and adjust them if necessary.