14
《
Operating procedure
》
1. Slide the power switch to the
【
MEAS
】
position and release it.
2. While in measurement mode, push SET for 3 seconds or longer. The display unit enters set up
mode and [F-1] is indicated.
3. Push
, and the setting contents of [F-1], [F-2], [F-3], [F-4], [F-5], [F-7], [F-10], [F-11], [F-12],
[F-13], [F-14] and [F-15] are indicated in sequence.
4. Values of the rated output (RO) and rated capacity (RC) indicated on the inspection sheet and
nameplate of the measuring device appear on the [F-3 ENROLMNT ADJ] display. Push SET
while confirming these are indicated.
5. [F-3] and the rated output (RO) will blink.
Enter the rated output (RO) indicated on the inspection sheet and nameplate.
Push
and select the digit to enter. Change the value by pushing
. Ensure the correct rated
output (RO) value is displayed, and push SET .
This completes the set up of the rated output (RO). Start set up of the rated capacity (RC) with
the following procedure.
6. [F-3] and the rated capacity (RC) will blink.
7. Enter the rated capacity (RC) (max.
〇〇
kN) indicated on the inspection sheet and nameplate.
Push
and select the digit to enter. Change the value by pushing . The number of digits to
be displayed during measurement should be considered.
Example: Measuring unit with 30kN rated capacity
To display without decimals
→
Enter [+00030]
To display to the 1st decimal place
→
Enter [+00300]
To display to the 2nd decimal place
→
Enter [+03000]
Ignore the decimal point when entering.
Ensure the correct rated capacity (RC) value is displayed, and push SET . This completes the
set up of the rated capacity (RC). Start set up of the decimal point with the following procedure.
8. After the decimal point blinks, push
and move the decimal point to the desired position.
Be sure to set to the decimal point position determined in the previous step.
Example Enter [+00030]
→
[+00030.] without decimal places
Enter [+00300]
→
[+0030.0]
Enter [+03000]
→
[+030.00]
9. Push SET and complete the position of the decimal point. If completed
correctly, [F-3] stops blinking.
If the input value is wrong, "CcErr" is displayed and "EA" appears at the
bottom left of the LCD. Push SET and go through the process again.
10. Push ZERO to complete the setting and return to the measurement mode.
EA
CAUTION
If there is an error in these set ups, correct measurement is not achieved.
All the previous set ups are not erased even when the power is turned off.
Therefore, resetting is not required unless another measuring device is used.
CcErr
kN
3. How to set up display unit
Summary of Contents for DynaForce
Page 5: ...kN 10kN 1020kgf oL oL Ad Ad 3 2 MEAS STRAIN HOLD 2 OFF 3 2 POWER...
Page 10: ...F 13 ENGLISH 4 1 F 14 3 4 F 14 11 F 11 TEDS F 12 4 1 3 F 15 4 1 F 10 4 1 EA...
Page 11: ...ATC 0 60 5 40 20 85 RH OFF 2 1 BIG...
Page 21: ...MEMO...
Page 22: ...MEMO...
Page 23: ...MEMO...
Page 24: ...No 0619 3 3 39 TEL 072 982 2312 FAX 072 980 2231 www big daishowa co jp 579 8013...