Operating procedure

1. Slide the power switch to the 

MEAS

 position and release it.

2. While in measurement mode, push  SET  for 3 seconds or longer. The display unit enters set up 

mode and [F-1] is indicated.

3. Push     

  

, and the setting contents of [F-1], [F-2], [F-3], [F-4], [F-5], [F-7], [F-10], [F-11], [F-12], 

[F-13], [F-14] and [F-15] are indicated in sequence.

4. Values of the rated output (RO) and rated capacity (RC) indicated on the inspection sheet and 

nameplate of the measuring device appear on the [F-3 ENROLMNT ADJ] display. Push  SET 
while confirming these are indicated.

5. [F-3] and the rated output (RO) will blink. 

Enter the rated output (RO) indicated on the inspection sheet and nameplate. 
Push  

     

 and select the digit to enter. Change the value by pushing    

  

. Ensure the correct rated 

output (RO) value is displayed, and push  SET . 
This completes the set up of the rated output (RO). Start set up of the rated capacity (RC) with 
the following procedure.

6. [F-3] and the rated capacity (RC) will blink.
7. Enter the rated capacity (RC) (max.

 〇〇

kN) indicated on the inspection sheet and nameplate.  

Push     

  

 and select the digit to enter. Change the value by pushing      . The number of digits to 

be displayed during measurement should be considered.

    Example: Measuring unit with 30kN rated capacity
      

  

To display without decimals 

 Enter [+00030]

      

  

To display to the 1st decimal place 

 Enter [+00300]

     

  

 To display to the 2nd decimal place 

 Enter [+03000]

        

Ignore the decimal point when entering.

Ensure the correct rated capacity (RC) value is displayed, and push  SET . This completes the 
set up of the rated capacity (RC). Start set up of the decimal point with the following procedure.

8. After the decimal point blinks, push  

    

 and move the decimal point to the desired position.  

Be sure to set to the decimal point position determined in the previous step.

    Example   Enter [+00030] 

 [+00030.] without decimal places

                  

Enter [+00300] 

 [+0030.0]

                  

Enter [+03000] 

 [+030.00]

9. Push  SET  and complete the position of the decimal point. If completed 

correctly, [F-3] stops blinking. 
If the input value is wrong, "CcErr" is displayed and "EA" appears at the 
bottom left of the LCD. Push  SET  and go through the process again.

10. Push  ZERO  to complete the setting and return to the measurement mode.

EA

CAUTION

If there is an error in these set ups, correct measurement is not achieved.

All the previous set ups are not erased even when the power is turned off.  

Therefore, resetting is not required unless another measuring device is used.

CcErr

kN

3. How to set up display unit

Summary of Contents for DynaForce

Page 1: ...O SET kN TYPE DFA 1 TYPE DFA 1 It is necessary to calibrate each measuring device Select MEAS position and indicate kN on the display before starting measuring OPERATION MANUAL Read the operation manu...

Page 2: ...hold measurement 2 7 End of measurement 2 8 Replacement of batteries P10 P11 P11 P11 P11 P12 P12 P12 P12 3 How to set up display unit 3 1 Confirming sensitivity calibration 3 2 Sensitivity calibration...

Page 3: ...1 1 ZERO 2 1 1 2 3 4 5 6 7 STRAIN POWER STRAIN MEAS POWER STRAIN MEAS Ser No 1 2 5 7 3 4 TYPE DFA 1 6...

Page 4: ...2 1 3 MAS JIS DIN ISO CAT ANSI OFF POWER POWER STRAIN MEAS POWER STRAIN MEAS 2 MEAS 0 0 kN ZERO ZERO POWER POWER ZERO STRAIN STRAIN 3...

Page 5: ...kN 10kN 1020kgf oL oL Ad Ad 3 2 MEAS STRAIN HOLD 2 OFF 3 2 POWER...

Page 6: ...3 2 F 3 F 3 MAX 30kN Rated output 3292 RC RO RC 30kN RO 3292 10kN 1020kgf MEAS F 1 8 F 13 F 1 F 2 F 3 F 4 F 5 F 7 F 10 F 11 F 12 F 13 F 14 F 15 F 3 RO RC...

Page 7: ...AS F 1 F 1 F 2 F 3 F 4 F 5 F 7 F 10 F 11 F 12 F 13 F 14 F 15 F 3 RO RC F 3 RO RO RO RO RC F 3 RC RC MAX kN 30kN 00030 1 00300 2 03000 RC RC 00030 00030 00300 0030 0 03000 030 00 F 3 EA 10 CcErrkN EA O...

Page 8: ...5 F 14 0 99 0 99 00 MEAS F 1 F 1 F 2 F 3 F 4 F 5 F 7 F 10 F 11 F 12 F 13 F 14 F 15 F 5 F 5 00 99 MIN F 5 00 99 SEC F 5 MEAS F 1 F 1 F 2 F 3 F 4 F 5 F 7 F 10 F 11 F 12 F 13 F 14 F 15 F 14 F 14 00 30 F...

Page 9: ...F 1 F 1 4 1 F 2 F 2 3 2 F 3 F 3 3 2 F 3 F 4 4 1 16 1 0 F 5 0 3 3 F 5 0 MIN 10 SEC F 1 F 1 F 2 F 3 F 4 F 5 F 7 F 10 F 11 F 12 F 13 F 14 F 15 F F F 7 4 1...

Page 10: ...F 13 ENGLISH 4 1 F 14 3 4 F 14 11 F 11 TEDS F 12 4 1 3 F 15 4 1 F 10 4 1 EA...

Page 11: ...ATC 0 60 5 40 20 85 RH OFF 2 1 BIG...

Page 12: ...h as entering values and for zero reset during measurement Displays ZERO on the LCD HOLD button To be used for changing the number of digits moving the decimal point or selecting peak hold mode during...

Page 13: ...se STRAIN mode as it is for error diagnosis by manufacturer If operated in STRAIN mode resetting is required as indicated in 3 How to set up display unit If the battery symbol is blinking the battery...

Page 14: ...n Measured in STRAIN mode Max measured value is maintained and displayed 1 After zero resetting push HOLD HOLD is displayed on the LCD to show that peak hold measurement is in progress 2 Install the m...

Page 15: ...ues is kN eg 10kN 1020kgf Operation procedure 1 Slide the power switch to the MEAS position and release it 2 While in measurement mode push SET for 3 seconds or longer The display unit enters set up m...

Page 16: ...igits to be displayed during measurement should be considered Example Measuring unit with 30kN rated capacity To display without decimals Enter 00030 To display to the 1st decimal place Enter 00300 To...

Page 17: ...ts of F 1 F 2 F 3 F 4 F 5 F 7 F 10 F 11 F 12 F 13 F 14 and F 15 are indicated in sequence 4 Select F 5 LOW POWER and push SET 5 F 5 and the power off value will blink Push or to set the time in a rang...

Page 18: ...F 3 F 4 For manufacturer s use only In case the set up is changed refer to 4 1 Changing the setting mode and return to the initial value Initial value AVERAGE 16 MIN VAL 1 NEAR ZR 0 F 5 Low power This...

Page 19: ...kN Key lock If the key lock function is set to ON it will always start up with the key lock ON when the power is turned on and key mark will be displayed Lock key Locks SET allowing use of ZERO and H...

Page 20: ...ature 5 C to 40 C Operating humidity 20 to 85 RH Condensation should be avoided Never exceed the above ranges Be sure to turn off power after use Replace both 2 AA batteries at the same time Dyna Forc...

Page 21: ...MEMO...

Page 22: ...MEMO...

Page 23: ...MEMO...

Page 24: ...No 0619 3 3 39 TEL 072 982 2312 FAX 072 980 2231 www big daishowa co jp 579 8013...

Reviews: