ENG
11
D.
Fix the whiteboxes (1601015 and 1502569) to the dental unit using Velcros
(1502565).
E.
Connect the pneumatic circuit (1502566) to the delivery unit and to the motors
according to the “5.3.2 DMX3 with 1 MX2 motor diagram” on page 12.
Note:
Please use pneumatic pipes with an internal diameter of 2.8 mm.
F.
Connect the electrovalves (1502523) to the DMX3 Whitebox (1601015) (1).
G.
Connect the DMX3 Whitebox (1601015) to the motor hose following the color
code (2) on the connector.
H.
Connect the Power supply reset switch to the DMX3 Whitebox (1601015) (3).
I.
Connect the Apple USB Cable (not provided in the set) to the DMX3 Whitebox (1601015).
J.
Drill a hole of diameter 19 mm in the unit with the hole saw (3300409). Pass the
Apple USB cable through the opening on the dental unit. To protect and avoid any
USB cable movement:
• fix it inside the unit with the provided cable zip tie (1) and its Self-adhesive
cable tie mount (1502535) (2),
• protect the opening with cable pass-through protection.
K.
Connect the Lightning cable to the Apple device.
L.
Connect the power supply cable.
M.
Configure the board through the App according to the iOptima application set-
tings chapter.
FIG. 7
FIG. 8
FIG. 9
FIG. 10
1
3
2
1
2