14
5.3.3 DMX3 with 2 MX2 motors procedure
A.
Drill a hole of diameter 12 mm in the unit to install the Power supply reset
switch (1502568).
B.
Connect together: Power supply reset switch (1502568), Power Supply PMP90
(1500666) (1) and Power Outlet Cable (1300067).
C.
Drill four holes of diameter 3.5 mm in the unit to fix the electrovalves (1502528)
to the dental unit using screws and nuts (1502574).
D.
Fix the whiteboxes (1601015 and 1502569) to the dental unit using Velcros
(1502572).
E.
Connect the pneumatic circuit (1502573) to the delivery unit and to the motors
according to the “5.3.4 DMX3 with 2 MX2 motors diagram” on page 16.
Note:
Please use pneumatic pipes with an internal diameter of 2.8mm.
F.
Connect the air switches to the motor holders.
G.
Connect the air switches to the DMX3 board (1).
H.
Connect the electrovalves (1502528) to the dual motor whitebox (1502569) (2).
FIG. 11
FIG. 12
FIG. 13
FIG. 14
2
1
2