6.6 Packing and storage
The device must be stored inside the
sterilization pouch in a dry and dust
free environment. If The temperature
must not exceed 55°C (131°F). If the
device will not be used for 7 days or
more after the sterilization, extract the
device from the sterilization pouch and
store it in the original package. If the
device is not stored in a sterilization
pouch or if the pouch is no longer
sterile, clean, lubricate and sterilize the
device before using it.
CAUTION
If the medical device has been stored re-
frigerated, allow it to warm up to room
temperature prior to use.
CAUTION
Comply with the expiration date of the
sterilization pouch which depends on
the storage conditions and the type of
packaging.
6.7 Servicing
Never dismantle the device. For all ser-
vicing or repair operations, you are
advised to contact your usual supplier
or Bien-Air Dental SA directly.
Note :The microsaw should be returned
for servicing at least one time per two
years due to possible wear. This op-
eration is strongly recommended for
maintaining the device in optimal work-
ing condition.
Note : It is recommended that the ir-
rigation ring be replaced regularly, at
least after every 50 patients treated.
7 Transport
& disposal
7.1 Transport
There are no particular transport and
storage conditions required.
7.2 Disposal
The disposal and/or recycling of ma-
terials
must
be
performed
in
accordance with the legislation in force.
The device must be recycled. In order to
avoid any risk of contamination, the
user must return it sterilized to his
dealer or contact an authorized body
for the treatment and recovery of this
type of equipment.