![Bien Air 1601098-001 Instructions For Use Manual Download Page 14](http://html1.mh-extra.com/html/bien-air/1601098-001/1601098-001_instructions-for-use-manual_2749371014.webp)
6.1.3 Microsaw and irrigation ring
l
For cleaning, disinfection, and ster-
ilization, if present, the irrigation
ring may be either cleaned sep-
arately or left on the microsaw.
For the last case, after a maximal
amount of 25 uses, the irrigation
ring
must
be
disconnected,
checked, and the microsaw with
the
irrigation
ring
must
be
cleaned, disinfected, and sterilized
separately.
WARNING
At any time, if the irrigation ring is in a
clogged state or damaged, proceed to
change with a new one.
6.2 Cleaning
Preparation
1. Disconnect the microsaw from the
motor.
2. Remove the cutting tool (see
).
3. Disconnect the irrigation line from
the microsaw (see
).
4. If present, the irrigation ring may be
either cleaned separately or left on
the microsaw.
5. After selecting the appropriate
nozzle, perform preliminary cleaning
of the microsaw by using the product
Aquacare. Spray the inside and the
outside of the device and inside the ir-
rigation tube.
6. Check that the coolant outlets are
rinsed out.
7. Dry the microsaw with non-woven
compress.
Internal and external cleaning
1. After selecting the appropriate
nozzle, spray the inside and the out-
side of the device as well as the
inside of the clamping ring with the
product Spraynet®.
2. Dry with non-woven compresses.
15