background image

9

5. Technical Specification

5.1 Product specifications

Sizes

GPro P1 Gutta Percha Obturation Pen

23.8mm×158.3mm×23.8mm

Charging base

75.5mm×149.7mm×62.6mm

Weight

GPro P1 Gutta Percha Obturation Pen

80g

Charging base

207g

Power adapter

167g

5.2 Technical parameters

Classification

Class II (AC/DC power adapter)

Optional preset temperatures

150°C-180°C-200°C-230°C

Time consumption for charging

About 2.5h (First charging needs 3 h)

Power supply

Input

AC100V-240V  50/60Hz 

800mA

Output

DC15V/1.6A

Battery capacity

Chargeable battery

2000mAh

5.3 Environmental parameters

Working

Storage

Temperature

+5°C~+40°C

-20°C~ +55°C

Humidity

30%~75%

10% ~ 93%

Air pressure

70kPa~106kPa

70kPa ~ 106kP

Warning:

Do not transport and store this device with hazardous articles such as combustible poisonous, 

caustic, or explosive.

Summary of Contents for GPRO P1

Page 1: ...Gutta Percha Obturation System Instruction for Use...

Page 2: ...k involved safety requirements are classed under the following indications Danger Always refer to personal injury Warning Refer to possible damage to property Table of Contents 1 Introduction 3 2 Inst...

Page 3: ...nded Use It is used to heat cut soften and downpack gutta percha to obturate root canals in endodontic treatment 1 3 Contraindications Patient allergic to known natural latex and metals such as stainl...

Page 4: ...And the design of left and right sides are perfectly symmetrical which enables either left hand or right hand operation 1 ON OFF button 2 Heating button 3 Heat insulation sleeve 4 Tip 5 Connecting ho...

Page 5: ...e first use Warning Do not drain battery Always charge to full to maintain battery life Recharge battery if the device has not been used for more than 1 month 2 Installation 2 1 Charge device 1 Instal...

Page 6: ...nto heating chamber completely Note To remove the tip hold near the turn of the tip and gently pull out Warning Needle remains hot after use Wait till display 0 C Warning Before each use clean disinfe...

Page 7: ...release on off button to select the temperature setting Temperature setting cycles as shown below Note The display shows 0 C when no button is pressed If the tip heats up or cools down the display sho...

Page 8: ...oint 4 7 mm short of the working length Move and set the endo stop accordingly 6 Insert the selected master Gutta Percha cone that is coated with sealer 7 Insert GPro P1 tip in the orifice of the cana...

Page 9: ...ower adapter Optional preset temperatures 150 C 180 C 200 C 230 C Time consumption for charging About 2 5h First charging needs 3 h Power supply Input AC100V 240V 50 60Hz 800mA Output DC15V 1 6A Batte...

Page 10: ...tch Ordinary equipment Used indoor only Device complies with WEEE directive Can be autoclaved Keep dry Handle with care Recovery Attention Please refer to the accompanying documents Humidity limit for...

Page 11: ...should be equipped with adequate protection such as goggles mask etc to prevent cross infection 7 The product should be in strict accordance with relevant operation specifications of medical authority...

Page 12: ...le and diffluent solvents for cleaning which will damage the surface of the device or cause the markings on the machine to fade 8 2 Daily maintenance 1 When the device is not used please turn off the...

Page 13: ...s the interior of the heating chamber is fixed by the cooling of the gutta percha 1 Power on and set the temperature to 200 C After the temperature reaches the set value pull out the pushing ram back...

Page 14: ...only to the first retail purchaser of a product and is not transferable or assignable Replacement components or products may be used and or refurbished components or products provided they are of lik...

Page 15: ...arranty is the repair or replacement of defective parts Beyes shall not be liable for and hereby disclaims any direct special indirect incidental exemplary or consequential damages or delays including...

Page 16: ...aplein 10 Ie Verd 2595AA The Hague Netherlands Tel 31645171879 English 31626669008 Dutch Email info beyes ca Website www beyes ca Printed in Canada ENI036 Rev 1 12 03 21 Federal law restricts this dev...

Reviews: