10. Troubleshooting

Fault

Cause

Solution

No 

indications, 

no response

1. Inadequate battery power

2. Battery is damaged.

3. The charging interface is short-

circuited, causing the lithium battery to 

enter a protection state;

4. GPro P1 Gutta Percha Obturation 

Pen is damaged.

1. Connect to power supply to 

charge. / Replace the battery.

2. Replace the battery.

3. Remove the substance that 

causes the short circuit, put the 

device into the charging base

to charge, and then the device will 

return to normal;

4. Contact local distributor or 

manufacturer.

Automatic 

shutdown

If there is no operation for 10 minutes, 

the device will automatic powers off.

Reboot

Tip works 

abnormally

1. The tip is damaged.

2. Malfunction of main unit

1. Replace the tip

2. Send it to the repair center.

The pushing 

ram cannot 

be pulled out

The portion of the pushing ram that 

enters the interior of the heating 

chamber is fixed by the cooling of the 

gutta-percha.

1. Power on and set the 

temperature to 200 °C. After the 

temperature reaches the set value, 

pull out the pushing ram back;

2. Contact your local dealer or our 

company.

Charging

failure after

connecting to 

power supply.

1. The power supply is not correctly 

connected;

2. The power supply is damaged, or 

the specification doesn’t match.

3. There are impurities on the contact 

thimble of charging base.

1. Unplug and reconnect.

2. Replace the battery.

3. Wipe the thimble with alcohol, 

dry it, and reconnect.

The service 

time after 

each 

charging is 

shortened

The battery ages and the battery 

capacity become smaller.

Contact local distributor or 

manufacturer to

buy new batteries for replacement.

OPEN code 

appears 

on display 

screen

1. The tip is damaged.

2. The tip is not installed.

3. The tip is not well installed.

1. Replace the tip.

2. Install the tip.

3. Unplug the tip, and reconnect.

If the problem still cannot be solved, please contact your local dealer or Beyes Dental Canada 

Inc. 

Summary of Contents for GPRO P1

Page 1: ...Gutta Percha Obturation System Instruction for Use...

Page 2: ...k involved safety requirements are classed under the following indications Danger Always refer to personal injury Warning Refer to possible damage to property Table of Contents 1 Introduction 3 2 Inst...

Page 3: ...nded Use It is used to heat cut soften and downpack gutta percha to obturate root canals in endodontic treatment 1 3 Contraindications Patient allergic to known natural latex and metals such as stainl...

Page 4: ...And the design of left and right sides are perfectly symmetrical which enables either left hand or right hand operation 1 ON OFF button 2 Heating button 3 Heat insulation sleeve 4 Tip 5 Connecting ho...

Page 5: ...e first use Warning Do not drain battery Always charge to full to maintain battery life Recharge battery if the device has not been used for more than 1 month 2 Installation 2 1 Charge device 1 Instal...

Page 6: ...nto heating chamber completely Note To remove the tip hold near the turn of the tip and gently pull out Warning Needle remains hot after use Wait till display 0 C Warning Before each use clean disinfe...

Page 7: ...release on off button to select the temperature setting Temperature setting cycles as shown below Note The display shows 0 C when no button is pressed If the tip heats up or cools down the display sho...

Page 8: ...oint 4 7 mm short of the working length Move and set the endo stop accordingly 6 Insert the selected master Gutta Percha cone that is coated with sealer 7 Insert GPro P1 tip in the orifice of the cana...

Page 9: ...ower adapter Optional preset temperatures 150 C 180 C 200 C 230 C Time consumption for charging About 2 5h First charging needs 3 h Power supply Input AC100V 240V 50 60Hz 800mA Output DC15V 1 6A Batte...

Page 10: ...tch Ordinary equipment Used indoor only Device complies with WEEE directive Can be autoclaved Keep dry Handle with care Recovery Attention Please refer to the accompanying documents Humidity limit for...

Page 11: ...should be equipped with adequate protection such as goggles mask etc to prevent cross infection 7 The product should be in strict accordance with relevant operation specifications of medical authority...

Page 12: ...le and diffluent solvents for cleaning which will damage the surface of the device or cause the markings on the machine to fade 8 2 Daily maintenance 1 When the device is not used please turn off the...

Page 13: ...s the interior of the heating chamber is fixed by the cooling of the gutta percha 1 Power on and set the temperature to 200 C After the temperature reaches the set value pull out the pushing ram back...

Page 14: ...only to the first retail purchaser of a product and is not transferable or assignable Replacement components or products may be used and or refurbished components or products provided they are of lik...

Page 15: ...arranty is the repair or replacement of defective parts Beyes shall not be liable for and hereby disclaims any direct special indirect incidental exemplary or consequential damages or delays including...

Page 16: ...aplein 10 Ie Verd 2595AA The Hague Netherlands Tel 31645171879 English 31626669008 Dutch Email info beyes ca Website www beyes ca Printed in Canada ENI036 Rev 1 12 03 21 Federal law restricts this dev...

Reviews: