BEYES Comfort-Sonic P6 Plus Instructions For Use Manual Download Page 8

8

• 

Water volume adjust: Step on the foot switch, and the tip will begin to vibrate, then turn the 

water control switch to form fine spray to cool down the handpiece and clean the teeth.

• 

The handpiece can be handled in the same gesture as a pen in hand.

• 

During the clinical treatment, be sure not to make the end of tip touch the teeth vertically 
and not to make the tip overexert on the surface of the teeth in case of hurting the teeth and 
damaging the tip

•  After finishing operation, keep the machine working for 30 seconds on the water supply 

condition in order to clean the handpiece and the scaling tip.

• 

Unscrew the scaling tip and pull out handpiece, then sterilize them.

Notice

Don’t pull the handpiece out when the foot switch is stepped on and the ultrasonic vibration is 
produced from the machine.

2.1.2 Instruction for main components of detachable handpiece (See picture 4):

• 

Nipple: The nipple can be removed. You can screw out the nipple and clean the pole with 
alcohol termly.

• 

Handpiece seal: The seal can’t be removed.

• 

Handpiece: The main part of the whole handpiece, can be autoclaved under the high 
temperature and pressure.

• 

The connector of the cable: Connect the handpiece with the water source and power supply of 
the main unit.

•  LED lamp, Light pipe: Clean them with purified water and sterilize them under the high 

temperature of 135

 o

C and high pressure of 0.22Mpa.

Notice

Keep dry when the detachable handpiece connects to the connector of the cable.

2.1.3 Instruction of torque wrench (See picture 5)

• 

The torque wrench’s structure is designed in special way which can control the strength of 
the scaling tip’s installation properly and correctly. It also can guarantee the operator screw or 
unscrew the scaling tip effectively and keep their hands away from being scratched.

• 

Operation

 

a. Take the scaling tip into the torque wrench as showed in picture 5.

 

b. Tip installation: Hold the handpiece, turn the tip toward direction as showed in figure 7 with  

 

the torque wrench. Turn twice more circles when the tip stops, then the tip will be installed.

 

c. Tip’s uninstallation: Hold the handpiece, turn the wrench in anti-clockwise direction direction.

 

d. Serilize it in sterilizer after each treatment.

 

e. The torque wrench must be cooled down naturally after sterilization to avoid scald when  

 

using for the next time.

 

f. Keep the torque wrench in a cool, dry and ventilated place and keep it clean.

Summary of Contents for Comfort-Sonic P6 Plus

Page 1: ...Comfort Sonic Ultrasonic Scaler P6 Plus Instruction for Use...

Page 2: ...3 2 Sterilization of scaling tips and endochuck P10 3 3 Serilization of torque wrench and edo wrench P10 3 4 Cleaning of tips endochuck torque wrench and endo wrench P10 3 5 Sterilization and cleanin...

Page 3: ...for scaling These features make COMFORT SONIC become a new generation product in the world dental market 1 2 Components 1 2 1 The components of the machine are listed in the packing list The scaling...

Page 4: ...gree against water used on the foot switch IPX1 Degree of safety of application in the presence of a Flammable Anesthetic Mixture with air or with Oxygen or Nitrous Oxide Equipment is not suitable for...

Page 5: ...icture 2 picture 2 picture 1 b Socket for foot switch Socket for power supply Water entrance Power switch Water control knob Plug of foot switch Foot switch socket Power supply Power cord plug Power c...

Page 6: ...chable handpiece are showed in picture 4 e Sketch map for how to install tip and endpchuck with wrench are showed in picture 5 LED lamp Nipple Light pipe Handpiece decorative ring The connector of cab...

Page 7: ...ugh to the power See picture 2 Switch on the main unit then the scaling indicator and the leading five lights of power regulator shine Step on the foot switch the tip begins to vibrate and the LED lam...

Page 8: ...the high temperature and pressure The connector of the cable Connect the handpiece with the water source and power supply of the main unit LED lamp Light pipe Clean them with purified water and steril...

Page 9: ...responsible for any damage of the torque wrench directly or indirectly made by any way in the above items 3 Sterilization and maintenance 3 1 Sterilization of detachable handpiece 3 1 1 Autoclaved in...

Page 10: ...Clean the water pipe by multi function syringe note 2 3 6 1 Troubleshooting The following sterilizing methods are forbiden a Putting handpiece into any liquid for boiling b Dipping handpiece in disinf...

Page 11: ...nge to a new O ring The ultrasonic doesn t vibrate The screw is loosen Tighten it Endochuck is damaged Change to a new endochunck LED light doesn t work Poor contact Contact tightly Something wrong wi...

Page 12: ...he socket easy to pull out to make sure it can be pull out in emergency 4 1 14 The device can only be equipped with the special power supply of original manufacturer 4 1 15 The power supply is NOT wat...

Page 13: ...don t invert it 4 4 2 Don t put it together with dangerous goods during transportation 4 4 3 Avoid solarization and getting wet in rain and snow during transportation 4 5 Working condition Enironment...

Page 14: ...h upon examination by Beyes prove to be defective This warranty extends only to the first retail purchaser of a product and is not transferable or assignable Replacement components or products may be...

Page 15: ...zed to sell the product at issue in the geographic territory where the purchaser is located or is not authorized to sell the product at issue within the medical animal health or dental market as the c...

Page 16: ...of the states in which he or she practices to use or order the use of this device 0197 Beyes Dental Canada Inc 23 595 Middlefield Road Toronto Ontario M1V 3S2 Canada Tel 1 855 603 1888 Fax 1 855 720 1...

Reviews: