There is still water flowing 

out after the power is off

There is impurity in the electric-

magnetic valve

Contact our dealers or us

The handpiece generates 

heat

The water control switch is in a low 

setting

Turn the water control switch to a 

higher grade[note2]

The amount of spouting 

water is too little

The water pressure is not high 

enough

Make the water pressure higher

The water pipe is blocked

Clean the water pipe by multi-function 

syringe[note2]

The vibration of tip becomes 

weak

The tip hasn’t been screwed on to the 

handpiece tightly

Screw the tip on the handpiece tightly 

(see picture 5)

The tip is loosen because of vibration

Screw on the tip tightly (see picture 5)

The coupling between the handpiece 

and the cable isn’t dry

Dry it by the hot air

The tip is damaged[note 3]

Change to a new tip

There is water seeping from 

the coupling between the 

handpiece and the cable

The waterproof “O” ring is damaged

Change to a new “O” ring

The ultrasonic doesn’t 

vibrate

The screw is loosen

Tighten it

Endochuck is damaged

Change to a new endochunck

LED light doesn't work

Poor contact

Contact tightly

Something wrong with LED light

Change a new one

LED lamp installed backwards

Please install the "+" of the LED lamp 

to the "+" of the handpiece

There is noise coming from 

the endochuck

The screw cap is loosen

Tighten it

If the problem still can’t be solved, please contact with local dealer or manufacturer.

3.6.2 Notes

• 

[Note 1] Turn the water control switch can adjust the water volume according to the symbol.

• 

[Note 2] Clean the water pipe with the Multi-function syringe of the dental unit (see picture 6)

 

 
 

a. Cut the water pipe from a distance of 10cm to 20cm from the water entrance.

 

b. Turn on the electricity and get through to the power.

 

c. Connect the multi-function syringe of dental unit to the water pipe.

 

d. Disassemble the tip or handpiece

 

e. Step on the foot switch.

 

f. Turn on the switch of the multi-function syringe, press the water into the machine and the  

 

impurity blocked in the water pipe can be eliminated.

Multi-function syringe

picture 6

Summary of Contents for Comfort-Sonic P6 Plus

Page 1: ...Comfort Sonic Ultrasonic Scaler P6 Plus Instruction for Use...

Page 2: ...3 2 Sterilization of scaling tips and endochuck P10 3 3 Serilization of torque wrench and edo wrench P10 3 4 Cleaning of tips endochuck torque wrench and endo wrench P10 3 5 Sterilization and cleanin...

Page 3: ...for scaling These features make COMFORT SONIC become a new generation product in the world dental market 1 2 Components 1 2 1 The components of the machine are listed in the packing list The scaling...

Page 4: ...gree against water used on the foot switch IPX1 Degree of safety of application in the presence of a Flammable Anesthetic Mixture with air or with Oxygen or Nitrous Oxide Equipment is not suitable for...

Page 5: ...icture 2 picture 2 picture 1 b Socket for foot switch Socket for power supply Water entrance Power switch Water control knob Plug of foot switch Foot switch socket Power supply Power cord plug Power c...

Page 6: ...chable handpiece are showed in picture 4 e Sketch map for how to install tip and endpchuck with wrench are showed in picture 5 LED lamp Nipple Light pipe Handpiece decorative ring The connector of cab...

Page 7: ...ugh to the power See picture 2 Switch on the main unit then the scaling indicator and the leading five lights of power regulator shine Step on the foot switch the tip begins to vibrate and the LED lam...

Page 8: ...the high temperature and pressure The connector of the cable Connect the handpiece with the water source and power supply of the main unit LED lamp Light pipe Clean them with purified water and steril...

Page 9: ...responsible for any damage of the torque wrench directly or indirectly made by any way in the above items 3 Sterilization and maintenance 3 1 Sterilization of detachable handpiece 3 1 1 Autoclaved in...

Page 10: ...Clean the water pipe by multi function syringe note 2 3 6 1 Troubleshooting The following sterilizing methods are forbiden a Putting handpiece into any liquid for boiling b Dipping handpiece in disinf...

Page 11: ...nge to a new O ring The ultrasonic doesn t vibrate The screw is loosen Tighten it Endochuck is damaged Change to a new endochunck LED light doesn t work Poor contact Contact tightly Something wrong wi...

Page 12: ...he socket easy to pull out to make sure it can be pull out in emergency 4 1 14 The device can only be equipped with the special power supply of original manufacturer 4 1 15 The power supply is NOT wat...

Page 13: ...don t invert it 4 4 2 Don t put it together with dangerous goods during transportation 4 4 3 Avoid solarization and getting wet in rain and snow during transportation 4 5 Working condition Enironment...

Page 14: ...h upon examination by Beyes prove to be defective This warranty extends only to the first retail purchaser of a product and is not transferable or assignable Replacement components or products may be...

Page 15: ...zed to sell the product at issue in the geographic territory where the purchaser is located or is not authorized to sell the product at issue within the medical animal health or dental market as the c...

Page 16: ...of the states in which he or she practices to use or order the use of this device 0197 Beyes Dental Canada Inc 23 595 Middlefield Road Toronto Ontario M1V 3S2 Canada Tel 1 855 603 1888 Fax 1 855 720 1...

Reviews: