33
Guidance and manufacturer’s declaration – electromagnetic emissions
The ApexPilot All-In-One
is intended for use in the electromagnetic environment specified below.
The customer or the user of the ApexPilot All-In-One should assure that is used in such an
environment.
Immunity test
IEC60601 test level
Compliance Level
Electromagnetic
environment - guide
Electrostatic
discharge(ESD)
EN 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least
30%.
Electrical fast
transient/burst,
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
Output lines
± 2 kV for power
supply lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
± 1 kV differential
mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5 % UT
( > 95 % dip in UT)
for 0.5 cycles
40 % UT
( 60 % dip in UT)
for 5 cycles
< 5 % UT
70 % UT
( 30 % dip in UT)
for 25 cycles
< 5 % UT
< 5 % UT
( > 95 % dip in UT)
for 5 s
< 5 % UT
( > 95 % dip in UT)
for 0.5 cycles
40 % UT
( 60 % dip in UT)
for 5 cycles
< 5 % UT
70 % UT
( 30 % dip in UT)
for 25 cycles
< 5 % UT
< 5 % UT
( > 95 % dip in UT)
for 5 s
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
ApexPilot All-In-One
requires continued
operation during power
mains interruptions,
it is recommended
that the ApexPilot
All-In-One be powered
from an uninterruptible
power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE:
UT is the a.c. mains voltage prior to application of the test level.