INTEGRATED ELECTRICAL CABINET

 

Installation, Operation, and Service Manual

 

 

Copyright

 2016:  

Bevco Engineering Company, Inc., 

W222N5739 Miller Way,  Sussex, WI 53089 USA 

IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems 

                        Page 9 of 14 

 

5.0  INSTALLATION 

 

  The enclosure must be mounted at a height so the top of the main circuit breaker 

handle, when in the ON position, does not exceed 6ft-7in. (2m) above the floor.  (Per 
National Electric Code (NEC # 404.8). 

  Incoming power is connected to the main circuit breaker located at the upper right of 

the Main Disconnect Panel.  Refer to the schematic drawing for the incoming conductor 
size range and tightening torques.  

  There are no conduit knockouts provided in the panel.  The installer must punch the 

conduit holes in the enclosure walls in the desired locations.  The conduits may enter 
on the top, bottom, either side, or in designated areas from the back of the enclosure.  

NOTE:

  When drilling or punching the conduit entry holes, protect the internal 

components from the falling metal chips. 

  All conductors must be sized per the wiring shown on the installation drawings or in the 

system configuration diagram. 

 

1.  Determine the system configuration.   

2.  Note the model of X-Ray Generator.  It will have a rating of 65kW or 80kW. 

3.  Connect the incoming power to the Line side of the Main CB, Q1, terminals L1, L2, and 

L3. 

4.  Connect the X-Ray Generator directly to the load side of contactor K2, terminals 2T1, 

4T2 and 6T3.   

5.  Connect the System Cabinet SC1 directly to the load side of contactor K6, terminals 

2T1, 4T2 and 6T3. 

6.  Connect the IEC Emergency Power Off (EPO) circuits to terminals 1 and 2 of X02. All 

EPOs should have normally closed contacts wired in series.  See EPO wiring diagram. 

7.  Room light controls are on terminals X05 1 through 4. Verify first if the Siemens System 

Cabinet is providing a dry contact or 24V AC or DC.  If 24V AC or DC is provided, 
connect the two wires coming from the Siemens System Cabinet to terminals 1 and 2 of 
X05.  If a dry contact is provided, connect the two wires coming from the Siemens 
System Cabinet to terminals 3 and 4 of X05. See Lighting Wiring Diagram. 

8.  Connect Room light and lighting panel to terminals 5 and 6 of X05. See Lighting Wiring 

Diagram. 

9. 

“X-Ray On” controls are on terminals X06 1 through 4. Verify first if the Siemens 
System Cabinet is providing a dry contact or 24V AC or DC.  If 24V AC or DC is 
provided, connect the two wires coming from the Siemens System Cabinet to terminals 
1 and 2 of X06. If a dry contact is provided, connect the two wires coming from the 
Siemens System Cabinet to terminals 3 and 4 of X06. See Lighting Wiring Diagram. 

10. Connect Warning light and lighting panel to terminals 5 and 6 of X06. See Lighting 

Wiring Diagram. 

11. Connect the Door Switch DSW1 to terminals 1 and 2 of X07.  See Door Switch Wiring 

Diagram. 

12. Connect the Door Switch Connections of the Siemens System Cabinet SC1 to the 

terminals 3 and 4 of X07.  See Door Switch Wiring Diagram. 

Summary of Contents for IECAX480V125A

Page 1: ...ications Installation Operation and Service Manual SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET for ARTIS Q Q ZEN ONE ARTIS ZEEGO ZEEGO ZEN ARTIS ZEE MULTIPURPOSE SINGLE PLANE IECAX480V125A Main Disconnect Panel ...

Page 2: ... of Bevco Engineering Company Troubleshooting and servicing should be performed by a qualified electrician De energize and lockout the main disconnect before servicing this panel Table of Contents 1 0 DESCRIPTION OF OPERATION 3 2 0 SYSTEM CONFIGURATION 4 3 0 SPECIFICATIONS 5 ELECTRICAL SPECIFICATIONS 5 SEISMIC SPECIFICATIONS 5 ENVIORNMENTAL SPECIFICATIONS 5 ELECTRICAL WIRING DIAGRAM 5 4 0 PANEL DI...

Page 3: ...3 and controlled by contactor K6 The contactor will open with any loss of power or by pressing any Emergency Power Off EPO pushbutton The control circuits for the EPOs are low voltage 24 VDC and are fully powered from within the panel The restart functionality and EPO circuitry is controlled with a safety relay K10 See page 11 for detail The white SAFETIES OK indicator light on the front of the pa...

Page 4: ...R K4 ROOM LIGHT RELAY K5 WARNING LIGHT RELAY EMERGENCY POWER OFF CIRCUITS EPO1 EPO2 LIGHTING POWER PANEL 120 277V KEY TO COMPONENTS Q1 MAIN CIRCUIT BREAKER Q2 GENERATOR 2 CIRCUIT BREAKER Q3 GENERATOR 1 CIRCUIT BREAKER F2 POWER SUPPLY FUSES F4 DC CIRCUIT BREAKER G01 DC POWER SUPPLY K1 SAFETY CONTACTOR K2 X RAY GENERATOR CONTACTOR K4 ROOM LIGHT RELAY K5 X RAY ON WARNING LIGHT RELAY K6 SYSTEM CABINET...

Page 5: ...nd Neutral 125 Amps cULus Short Circuit Current Rating is 35 000 Amperes RMS Symmetrical 480V SEISMIC SPECIFICATIONS This Integrated Electrical Cabinet has been certified by an independent California structural engineer in conformance with the shake testing requirements of ICC AC 156 The California OSHPD number is OSP 0457 10 The seismic performance characteristics are as follows SDS g 2 56 z h 1 ...

Page 6: ...el may be recessed into the wall up to 5 inches for semi flush installations See Illustrations below for typical front and side views with the mounting hole pattern Refer to the drawing set provided with the panel for total weight and center of gravity dimensions PANEL EXTERNAL PHYSICAL DIMENSIONS Parameter Dimensions Height 30 0 inches 762 mm Width 20 0 inches 508 mm Total Depth 9 44 inches 244 m...

Page 7: ... CABINET Installation Operation and Service Manual Copyright 2016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 7 of 14 CENTER OF GRAVITY ...

Page 8: ...INET Installation Operation and Service Manual Copyright 2016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 8 of 14 MOUNTING HOLE DIMENSIONS ...

Page 9: ...ectly to the load side of contactor K2 terminals 2T1 4T2 and 6T3 5 Connect the System Cabinet SC1 directly to the load side of contactor K6 terminals 2T1 4T2 and 6T3 6 Connect the IEC Emergency Power Off EPO circuits to terminals 1 and 2 of X02 All EPOs should have normally closed contacts wired in series See EPO wiring diagram 7 Room light controls are on terminals X05 1 through 4 Verify first if...

Page 10: ...e Manual Copyright 2016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 10 of 14 INSTALLATION continued TERMINAL LAYOUT Located in the bottom left of the Disconnect Panel EPO WIRING ...

Page 11: ...Installation Operation and Service Manual Copyright 2016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 11 of 14 LIGHTING WIRING DOOR SWITCH WIRING ...

Page 12: ...k the OUT light on the safety relay if not solid green press and hold the RESET button for five seconds then release The lights should flash and the DEVICE and IN light should be green Press the Start button contactor K2 should pull in If the DEVICE light is red the safety relay is defective and must be replaced For other Safety Relay Conditions refer to the diagram and chart next page PDA Off cir...

Page 13: ...016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 13 of 14 K10 Safety Relay configurations The Safety Relay DIP switches are set as follows from the factory 1 2 3 4 5 6 7 8 On On Off On On On On Off ...

Page 14: ...ervicing 8 0 REPLACEMENT PARTS 9 0 DOCUMENT REVISION HISTORY REV DATE AUTHOR DESCRIPTION OF CHANGE 1 10 03 2013 RJB Initial Release 2 02 27 2015 RJB Added Artis Tradenames on first page 3 03 09 2015 XYUE Update the manual based on schematic 50 03 30 16 DRO IEC Redesign 51 07 10 18 JT Update of system config data PARTS TECHNICAL SUPPORT AND SERVICE FOR INQUIRIES DURING INSTALLATION CONTACT BEVCO EN...

Reviews: