background image

 

INTEGRATED ELECTRICAL CABINET

 

Installation, Operation, and Service Manual

 

 

Copyright

 2016:  

Bevco Engineering Company, Inc., 

W222N5739 Miller Way,  Sussex, WI 53089 USA 

IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems 

                        Page 4 of 14 

 

2.0  SYSTEM CONFIGURATION 

 

UTILITY 

SUPPLY 

480V 3-PHASE 

100A MAX 

Q1 

125A 

WIRESIZE: 

6-3/0AWG

 

K2 

185A MAX 

WIRE SIZE: 

10-2/0AWG

 

 

F2 

4A 

 

 

G01 

DC POWER 

SUPPLY 

 

DC CONTROL CIRCUITS 

K1 – SAFETY CONTACTOR 

K4 – ROOM LIGHT RELAY 

K5 – WARNING LIGHT RELAY 

EMERGENCY POWER OFF 

CIRCUITS 

 

EPO1 

EPO2 

 

LIGHTING 

POWER PANEL 

120/277V 

 

KEY TO COMPONENTS: 
Q1 – MAIN CIRCUIT BREAKER 
Q2 – GENERATOR 2 CIRCUIT BREAKER 
Q3 – GENERATOR 1 CIRCUIT BREAKER 
F2 – POWER SUPPLY FUSES 
F4 – DC CIRCUIT BREAKER 
G01 – DC POWER SUPPLY 
K1 – SAFETY CONTACTOR 
K2 – X-RAY GENERATOR CONTACTOR 
K4 – ROOM LIGHT RELAY 
K5 – “X-RAY ON” WARNING LIGHT RELAY 
K6 – SYSTEM CABINET CONTACTOR 

 

GENERATOR 1 

100A MAX 

 

 

F4 

5A 

 

 

ROOM LIGHTS 

20A MAX 

 

“X-RAY ON” 

WARNING LIGHT 

20A MAX 

EPO3 

Q3 

ADJUSTABLE 

30A 

 

K6 

30A 

WIRE SIZE: 

14-8AWG 

 

Q2 

100A

 

SC1 

DOOR SWITCH 

CONTROL 

ROOM LIGHT CONTROL 

X-RAY ON CONTROL 

30A MAX 

 

Summary of Contents for IECAX480V125A

Page 1: ...ications Installation Operation and Service Manual SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET for ARTIS Q Q ZEN ONE ARTIS ZEEGO ZEEGO ZEN ARTIS ZEE MULTIPURPOSE SINGLE PLANE IECAX480V125A Main Disconnect Panel ...

Page 2: ... of Bevco Engineering Company Troubleshooting and servicing should be performed by a qualified electrician De energize and lockout the main disconnect before servicing this panel Table of Contents 1 0 DESCRIPTION OF OPERATION 3 2 0 SYSTEM CONFIGURATION 4 3 0 SPECIFICATIONS 5 ELECTRICAL SPECIFICATIONS 5 SEISMIC SPECIFICATIONS 5 ENVIORNMENTAL SPECIFICATIONS 5 ELECTRICAL WIRING DIAGRAM 5 4 0 PANEL DI...

Page 3: ...3 and controlled by contactor K6 The contactor will open with any loss of power or by pressing any Emergency Power Off EPO pushbutton The control circuits for the EPOs are low voltage 24 VDC and are fully powered from within the panel The restart functionality and EPO circuitry is controlled with a safety relay K10 See page 11 for detail The white SAFETIES OK indicator light on the front of the pa...

Page 4: ...R K4 ROOM LIGHT RELAY K5 WARNING LIGHT RELAY EMERGENCY POWER OFF CIRCUITS EPO1 EPO2 LIGHTING POWER PANEL 120 277V KEY TO COMPONENTS Q1 MAIN CIRCUIT BREAKER Q2 GENERATOR 2 CIRCUIT BREAKER Q3 GENERATOR 1 CIRCUIT BREAKER F2 POWER SUPPLY FUSES F4 DC CIRCUIT BREAKER G01 DC POWER SUPPLY K1 SAFETY CONTACTOR K2 X RAY GENERATOR CONTACTOR K4 ROOM LIGHT RELAY K5 X RAY ON WARNING LIGHT RELAY K6 SYSTEM CABINET...

Page 5: ...nd Neutral 125 Amps cULus Short Circuit Current Rating is 35 000 Amperes RMS Symmetrical 480V SEISMIC SPECIFICATIONS This Integrated Electrical Cabinet has been certified by an independent California structural engineer in conformance with the shake testing requirements of ICC AC 156 The California OSHPD number is OSP 0457 10 The seismic performance characteristics are as follows SDS g 2 56 z h 1 ...

Page 6: ...el may be recessed into the wall up to 5 inches for semi flush installations See Illustrations below for typical front and side views with the mounting hole pattern Refer to the drawing set provided with the panel for total weight and center of gravity dimensions PANEL EXTERNAL PHYSICAL DIMENSIONS Parameter Dimensions Height 30 0 inches 762 mm Width 20 0 inches 508 mm Total Depth 9 44 inches 244 m...

Page 7: ... CABINET Installation Operation and Service Manual Copyright 2016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 7 of 14 CENTER OF GRAVITY ...

Page 8: ...INET Installation Operation and Service Manual Copyright 2016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 8 of 14 MOUNTING HOLE DIMENSIONS ...

Page 9: ...ectly to the load side of contactor K2 terminals 2T1 4T2 and 6T3 5 Connect the System Cabinet SC1 directly to the load side of contactor K6 terminals 2T1 4T2 and 6T3 6 Connect the IEC Emergency Power Off EPO circuits to terminals 1 and 2 of X02 All EPOs should have normally closed contacts wired in series See EPO wiring diagram 7 Room light controls are on terminals X05 1 through 4 Verify first if...

Page 10: ...e Manual Copyright 2016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 10 of 14 INSTALLATION continued TERMINAL LAYOUT Located in the bottom left of the Disconnect Panel EPO WIRING ...

Page 11: ...Installation Operation and Service Manual Copyright 2016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 11 of 14 LIGHTING WIRING DOOR SWITCH WIRING ...

Page 12: ...k the OUT light on the safety relay if not solid green press and hold the RESET button for five seconds then release The lights should flash and the DEVICE and IN light should be green Press the Start button contactor K2 should pull in If the DEVICE light is red the safety relay is defective and must be replaced For other Safety Relay Conditions refer to the diagram and chart next page PDA Off cir...

Page 13: ...016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 13 of 14 K10 Safety Relay configurations The Safety Relay DIP switches are set as follows from the factory 1 2 3 4 5 6 7 8 On On Off On On On On Off ...

Page 14: ...ervicing 8 0 REPLACEMENT PARTS 9 0 DOCUMENT REVISION HISTORY REV DATE AUTHOR DESCRIPTION OF CHANGE 1 10 03 2013 RJB Initial Release 2 02 27 2015 RJB Added Artis Tradenames on first page 3 03 09 2015 XYUE Update the manual based on schematic 50 03 30 16 DRO IEC Redesign 51 07 10 18 JT Update of system config data PARTS TECHNICAL SUPPORT AND SERVICE FOR INQUIRIES DURING INSTALLATION CONTACT BEVCO EN...

Reviews: