Technical Description Concerning Electromagnetic Emission

Table 1: Declaration - electromagnetic emissions

Guidance and manufacturer’s declaration – electromagnetic immunity

The TL 70 is intended for use in the electromagnetic environment specified below.  

The customer or the user should assure that it is used in such an environment.

Immunity test

IEC 60601

test level

Compliance

level

Electromagnetic environment –

guidance

Electrostatic 

discharge (ESD) 

IEC 61000-4-2

±8 kV contact

±15 kV air

±8 kV contact

±15 kV air

Floors should be wood, concrete 

or ceramic tile. If floors are covered 

with synthetic material, the relative 

humidity should be at least 30%

Electrical fast 

transient/burst 

IEC 61000-4-4

±2 kV for power

supply lines

±1 kV for input/output 

lines

±2 kV for power

supply lines

Mains power quality should be that 

of a typical commercial or hospital 

environment.

Surge

IEC 61000-4-5

±1 kV line to line

±2 kV line to earth

±2 kV line to earth

Mains power quality should be that 

of a typical commercial or hospital 

environment.

Voltage dips, 

short interrup-

tions and vol-

tage variations 

on power supp-

ly input lines

IEC 61000-4-11

<5% U

T

 (>95% dip in U

T

for 0.5 cycle

40% U

T

 (60% dip in U

T

for 5 cycles

70% U

T

 (30% dip in U

T

for 25 cycles

<5% U

T

 (>95% dip in U

T

for 5 seconds

<5% U

T

 (>95% dip in U

T

for 0.5 cycle

40% U

T

 (60% dip in U

T

for 5 cycles

70% U

T

 (30% dip in U

T

for 25 cycles

<5% U

T

 (>95% dip in U

T

for 5 seconds

Mains power quality should be that 

of a typical commercial or hospital 

environment. If the user of the 

model TL 70 requires continued 

operation during power mains in-

terruptions, it is recommended that 

the TL 70 daylight therapy lamp is 

to be powered from an uninterrup-

tible power supply or a battery.

Power Frequen-

cy (50/60HZ) 

magnetic field 

IEC 61000-4-8

3 A/m

Not applicable

Not applicable

NOTE

 U

T

 is the a.c. mains voltage prior to application of the test level.

Summary of Contents for TL 70

Page 1: ...EN Daylight therapy lamp Instruction for use 2 ELECTROMAGNETIC COMPATIBILITY INFORMATION 9 TL 70 0598...

Page 2: ...ss This hormone tells the body when it is time to sleep In the months when there is less sunlight therefore production of melatonin increases This makes it harder to get up in the morning since the bo...

Page 3: ...ble to other users and observe the information they contain Safety notes Warning The light of the daylight therapy lamp is only intended for use on the human body Before use ensure that all packaging...

Page 4: ...unit can be dangerous and may cause injury or improper results The main material of the case is made of PC ABS Be cautious of potential allergic reactions to these ma terials General notes Important...

Page 5: ...t 6 Operation 1 Insert the mains part into the connector 2 Switch on the light Press the ON OFF button 3 LED display timer The daylight therapy lamp TL 70 indicates your current treatment time with th...

Page 6: ...n the distance Distance Duration Up to 10 cm hour day Up to 25 cm 1 hour day Up to 40 cm 2 hours day Basically the closer you are to the source of light the less time you should use it 7 Enjoying ligh...

Page 7: ...m Operating conditions 5 C to 35 C 15 90 relative humidity atmospheric pressure range 86 106 kpa Storage conditions 20 C to 50 C 15 90 relative humidity atmospheric pressure range 86 106 kpa Product c...

Page 8: ...first in stance see the attached International Service list of service addresses The buyer will then receive further information about the processing of the warranty claim e g where they can send the...

Page 9: ...t Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The models TL 70 use RF energy only for its internal function Therefore its RF emissions are very low and are not likely...

Page 10: ...cal commercial or hospital environment Surge IEC 61000 4 5 1 kV line to line 2 kV line to earth 2 kV line to earth Mains power quality should be that of a typical commercial or hospital environment Vo...

Page 11: ...trengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur In the vicinity of equip...

Page 12: ...ing a minimum distance between portable and mobile RF communications equipment transmitters and the model TL 70 as recommended below according to the maximum output power of the communica tions equipm...

Reviews: