8. Cleaning and disinfection

Atomiser and accessories

 Warning 

Observe the following hygiene instructions in order to 

avoid any health risks.

 The atomiser

 [7] and accessories are intended for 

multiple use. Please note that cleaning requirements 

differ according to the different application areas.

Notes:

• The atomiser and accessories must not be me-

chanically cleaned with brushes or similar as this 

can cause irreparable damage and the success of 

the treatment can no longer be guaranteed.

•  For additional requirements regarding the necessary 

hygienic preparations (hand washing, handling of 

medicines or inhalation solutions) in high-risk groups 

(e.g. cystic fibrosis patients), contact your doctor.

• Ensure that all parts are properly dried following all 

cleaning and disinfection. Residual moisture or water 

droplets can result in an increased risk of bacterial 

growth.

Preparation

• Directly following each treatment, all parts of the 

atomiser [7] and any accessories used must be 

cleaned of medicine residues and contamination.

• To do this, dismantle the atomiser [7] into its com-

ponent parts.

•  Remove the mouthpiece [10] from 

the atomiser.

•  Dismantle the atomiser by twisting 

the upper section anti-clockwise 

against the medicine container [9].

•  Remove the nozzle attachment [8] 

from the medicine container [9].

•  The nebuliser is later reassembled 

in the reverse order.

Cleaning

 Important

Before you clean the unit, always switch it off, unplug 

it and let it cool off.

The 

atomiser

 and the used 

accessories

 such as the 

mouthpiece, mask, nozzle attachment, etc. must be 

washed with hot but not boiling water after each use.

Carefully dry all components using a soft cloth. Once 

all components are fully dry, reassemble the atomiser 

and place the components in a dry, sealed container 

or carry out disinfection.
When cleaning, ensure that all residues are removed. 

Never use any substances that may be poisonous 

when in contact with the skin or mucous membranes, 

or when swallowed or inhaled.
Use a soft, dry cloth and non-abrasive cleaning prod-

ucts to clean the 

device

.

Do not use abrasive detergents and never immerse 

the unit in water.

 Important

• Ensure that no water penetrates inside the unit!

•  Do not put the device or accessories in a dishwasher.

• Do not touch the unit with wet hands while it is 

plugged in; do not allow any water to be sprayed 

onto the unit. The unit must be operated only when 

it is completely dry.

•  Do not spray any liquids into the ventilation slots. Any 

liquid that penetrates into the unit can damage the 

electrical parts or other components of the nebuliser 

and impair the function of the device.

Condensation, care of the hose

Depending on the environmental conditions, con-

densation may collect in the hose. To prevent against 

bacterial growth and to ensure effective therapeutic 

use, this moisture must always be removed. To remove 

moisture from the hose, proceed as follows:

• Remove the hose [6] from the atomiser [7].

• Leave the hose plugged into the nebuliser [4].

•  Run the nebuliser until all the moisture has been dis-

placed by the flow of air.

•  In the event of significant contamination, replace the 

hose.

Disinfection

Follow the steps listed below carefully in order to disin-

fect your atomiser and the accessories. It is advisable 

to disinfect the individual components after the last 

time they are used each day at the latest.

(For disinfection, you will simply require a small amount 

of clear vinegar and distilled water).

•  First clean the atomiser and the accessories as de-

scribed under “Cleaning”.

• Place the disassembled 

atomiser

 and 

mouthpiece

 

in boiling water for 5 minutes.

• For the remaining accessories, use a vinegar solution 

consisting of ¼ vinegar and ¾ distilled water. Make 

sure that the volume is sufficient to fully submerge 

the parts in the solution.

  Leave the components in the vinegar solution for 

30 minutes.

• Rinse the parts with water and dry carefully with a 

soft cloth.

8

9

Summary of Contents for IH 18

Page 1: ...IH 18 EN Nebuliser Instructions for use ...

Page 2: ...chi deep into the lungs Exhale normally Once prepared the device is suitable for reuse The prep aration includes the replacement of all accessory parts including atomiser and air filter as well as disinfection of the surfaces of the device with a standard disinfectant Please note that all accessories should be replaced if the device is used by more than one person We recommend that you replace the...

Page 3: ...ldren Do not use attachments that are not recommended by the manufacturer Connect the unit only to the mains voltage listed on the type plate Do not dip the unit into water and do not use it in wet rooms Do not allow any liquid to penetrate the unit Protect the unit from strong impacts Never touch the mains cable with wet hands this can cause an electric shock Do not pull the mains plug from the s...

Page 4: ...gainst dust dirt and moisture Never cover the unit during operation Do not use the unit in very dusty environments Switch the unit off immediately if it is defective or malfunctioning The manufacturer shall not be held liable for damage or injuries caused by improper or incorrect use Repairs Note Never open or attempt to repair the unit yourself as otherwise proper function is no longer guaranteed...

Page 5: ...t medicines are to be inhaled in suc cession during therapeutic use the atomiser 7 must be rinsed under warm tap water after each use For information about this see Cleaning and disinfec tion on page 7 Follow the instructions for changing the filter con tained in these instructions Before each use check that the hose connections on the nebuliser 4 and the atomiser 7 are tight Before use check that...

Page 6: ...g and drops for application to the skin or for use in steam baths are wholly unsuitable for inha lation using a nebuliser These substances are often highly viscous and can impair the correct function of the device and hence affect the effectiveness of its application in the long term In the case of oversensitivity of the bronchial system medicines containing essential oils can sometimes cause acut...

Page 7: ...ver use any substances that may be poisonous when in contact with the skin or mucous membranes or when swallowed or inhaled Use a soft dry cloth and non abrasive cleaning prod ucts to clean the device Do not use abrasive detergents and never immerse the unit in water Important Ensure that no water penetrates inside the unit Do not put the device or accessories in a dishwasher Do not touch the unit...

Page 8: ...ocal authorities 10 Troubleshooting Problem Ques tion Possible Cause Remedy The atomiser produces little or no aerosol 1 Too much or too little medicine in the atomiser Minimum 2 ml Maximum 8 ml 2 Check that the nozzles are not blocked If necessary clean the nozzle e g by rinsing Use the atomiser again CAUTION Only poke imple ments through the fine holes from the underside of the nozzle 3 The atom...

Page 9: ...060 hPa Aerosol Properties Flow 5 31 l min Aerosol delivery 0 171 ml Aerosol delivery rate 0 082 ml min Particle size MMAD 3 16 µm The serial number is located on the device or in the battery compartment Subject to technical modifications Particle size diagram Measurements were performed using a sodium fluoride solution with a Next Generation Impactor NGI This diagram may not be applicable for sus...

Page 10: ...er sonal purposes in the context of domestic use German law shall apply During the warranty period should this product prove to be incomplete or defective in functionality in accord ance with the following provisions Beurer shall carry out a repair or a replacement delivery free of charge in accordance with these warranty conditions If the buyer wishes to make a warranty claim they should approach...

Page 11: ...11 ...

Page 12: ...12 601 16_IH18_2019 12 16_05_IM2_BEU_EN Beurer GmbH Söflinger Straße 218 89077 Ulm Germany www beurer com www beurer healthguide com ...

Reviews: