Beurer IH 18 Instructions For Use Manual Download Page 5

5

5. Initial use

Setup

Remove the unit from the packaging. Place it on a flat 

surface.

Ensure that the ventilation slits are free.

Before using the unit for the first time

 Note

• The atomiser and accessories must be cleaned and 

disinfected before they are used for the first time. For 

information about this, see “Cleaning and disinfec-

tion” on page 7.

• Attach the compressed-air hose [6] at the bottom of 

the medicine container [9].

• Connect the other end of the hose [6] to the hose 

connector [4] on the nebuliser by lightly turning it.

Mains connection

Connect the unit only to the mains voltage listed on 

the type plate.

• Insert the mains cable plug [1] into a suitable socket.

• Push the plug completely into the mains socket.

 Note

• Ensure that there is a socket near to where the unit 

will stand.

• Lay the mains cable in such a way that no one can 

trip over it.

•  To disconnect the nebuliser from the mains after use, 

first switch off the device and then remove the plug 

from the socket.

6. Operation

 Caution

• For hygiene reasons, the atomiser [7] and the ac-

cessories must be cleaned after each treatment and 

disinfected after the last treatment of the day.

• The accessories may be used only by one person; 

use by several people is not recommended.

•  If several different medicines are to be inhaled in suc-

cession during therapeutic use, the atomiser [7] must 

be rinsed under warm tap water after each use. For 

information about this, see “Cleaning and disinfec-

tion” on page 7.

• Follow the instructions for changing the filter con-

tained in these instructions.

• Before each use, check that the hose connections 

on the nebuliser [4] and the atomiser [7] are tight.

• Before use, check that the device is functioning cor-

rectly. To do this, switch on the nebuliser (including 

the connected atomiser, but without any medication) 

for a short time. If the atomiser [7] expels air, the 

device is functioning correctly.

1. Inserting the nozzle attachment

•  To open the atomiser [7], twist the upper section anti-

clockwise against the medicine container [9]. Place 

the nozzle attachment [8] in the medicine container 

[9].

• Ensure that the cone for administering medication fits 

well on the cone for the air duct inside the atomiser 

[7].

2. Filling the atomiser 

•  Pour an isotonic saline solution or 

the medicine directly into the med-

icine container [9]. Do not overfill! 

The maximum recommended fill 

level is 6 ml !

•  Use medication only on the advice 

of a physician, and check the ap-

propriate inhalation duration and 

quantity for your needs.

• If the required volume of medi-

cation is less than 2 ml, top up this quantity to at 

least 4 ml using an isotonic saline solution. Viscous 

medicines may also require dilution. Always follow 

the instructions of your doctor.

8

9

Summary of Contents for IH 18

Page 1: ...IH 18 EN Nebuliser Instructions for use ...

Page 2: ...chi deep into the lungs Exhale normally Once prepared the device is suitable for reuse The prep aration includes the replacement of all accessory parts including atomiser and air filter as well as disinfection of the surfaces of the device with a standard disinfectant Please note that all accessories should be replaced if the device is used by more than one person We recommend that you replace the...

Page 3: ...ldren Do not use attachments that are not recommended by the manufacturer Connect the unit only to the mains voltage listed on the type plate Do not dip the unit into water and do not use it in wet rooms Do not allow any liquid to penetrate the unit Protect the unit from strong impacts Never touch the mains cable with wet hands this can cause an electric shock Do not pull the mains plug from the s...

Page 4: ...gainst dust dirt and moisture Never cover the unit during operation Do not use the unit in very dusty environments Switch the unit off immediately if it is defective or malfunctioning The manufacturer shall not be held liable for damage or injuries caused by improper or incorrect use Repairs Note Never open or attempt to repair the unit yourself as otherwise proper function is no longer guaranteed...

Page 5: ...t medicines are to be inhaled in suc cession during therapeutic use the atomiser 7 must be rinsed under warm tap water after each use For information about this see Cleaning and disinfec tion on page 7 Follow the instructions for changing the filter con tained in these instructions Before each use check that the hose connections on the nebuliser 4 and the atomiser 7 are tight Before use check that...

Page 6: ...g and drops for application to the skin or for use in steam baths are wholly unsuitable for inha lation using a nebuliser These substances are often highly viscous and can impair the correct function of the device and hence affect the effectiveness of its application in the long term In the case of oversensitivity of the bronchial system medicines containing essential oils can sometimes cause acut...

Page 7: ...ver use any substances that may be poisonous when in contact with the skin or mucous membranes or when swallowed or inhaled Use a soft dry cloth and non abrasive cleaning prod ucts to clean the device Do not use abrasive detergents and never immerse the unit in water Important Ensure that no water penetrates inside the unit Do not put the device or accessories in a dishwasher Do not touch the unit...

Page 8: ...ocal authorities 10 Troubleshooting Problem Ques tion Possible Cause Remedy The atomiser produces little or no aerosol 1 Too much or too little medicine in the atomiser Minimum 2 ml Maximum 8 ml 2 Check that the nozzles are not blocked If necessary clean the nozzle e g by rinsing Use the atomiser again CAUTION Only poke imple ments through the fine holes from the underside of the nozzle 3 The atom...

Page 9: ...060 hPa Aerosol Properties Flow 5 31 l min Aerosol delivery 0 171 ml Aerosol delivery rate 0 082 ml min Particle size MMAD 3 16 µm The serial number is located on the device or in the battery compartment Subject to technical modifications Particle size diagram Measurements were performed using a sodium fluoride solution with a Next Generation Impactor NGI This diagram may not be applicable for sus...

Page 10: ...er sonal purposes in the context of domestic use German law shall apply During the warranty period should this product prove to be incomplete or defective in functionality in accord ance with the following provisions Beurer shall carry out a repair or a replacement delivery free of charge in accordance with these warranty conditions If the buyer wishes to make a warranty claim they should approach...

Page 11: ...11 ...

Page 12: ...12 601 16_IH18_2019 12 16_05_IM2_BEU_EN Beurer GmbH Söflinger Straße 218 89077 Ulm Germany www beurer com www beurer healthguide com ...

Reviews: