• The codes below are printed on batteries
containing harmful substances:
Pb = Battery contains lead,
Cd = Battery contains cadmium,
Hg = Battery contains mercury.
12. Specifications
Device
Model no.
BM 81
Type
MD6000
Measurement method
Oscillometric, non-invasive blood pressure measurement on the
upper arm
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 50 – 250 mmHg,
diastolic 30 – 200 mmHg,
Pulse 40 –180 beats/minute
Display accuracy
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement inaccuracy
Max. permissible standard deviation according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 120 memory spaces
Dimensions
L 175 mm x W 117 mm x H 50 mm
Weight
Approx. 518 g (without batteries, with cuff)
Cuff size
24 to 40 cm
Permissible operating conditions
+5 °C to + 40 °C,
15 % – 90 % relative air humidity (non-condensing)
Permissible storage and transport
conditions
- 25 °C to + 70 °C,
up to 90%, non-condensing Atmospheric pressure: 700 hPa to 1060 hPa
Power supply
4 x 1,5 V
AAA batteries
Battery life
For approx. 120 measurements, depending on the blood pressure
level and/or pump pressure
Classification
Internal supply, IP22 , no AP or APG, continuous operation, applica
-
tion part type BF
Data transfer via
Bluetooth
®
wireless
technology
The blood pressure monitor uses
Bluetooth
®
low energy techno -
logy,
2402MHz – 2480MHz frequency band,
Transmission power max. 4 dBm,
Compatible with
Bluetooth
®
4.2 smartphones/tablets
The serial number is located on the device or in the battery compartment.
Technical information is subject to change without notification to allow for updates.
Service
life
10,000 measurement at least.