15
Select memory
You have two memories (50 memory spaces each) in order
to save the test results of 2 different people separately,
or else save measurements in the morning and evening
separately.
Select the required memory space by pressing the user
selection button
.
Measuring blood pressure
• Put on the cuff as described previously and assume the
position in which you want to carry out the measurement.
• Press the “User-Switching key”
to select a memo-
ry zone, and press the “
” key 2x to start measurement
in the chosen memory zone. After checking the display
with all digits lighting up, the monitor will automatically
inflate. Following the self-test, during which all display
elements briefly appear, the measurement begins. During
inflation the device already determines measured values
used to estimate the required inflation pressure. If this
pressure is insufficient, the device automatically inflates
by another 40 mmHg (Real Fuzzy Logic).
• Then the pressure in the cuff is slowly released and the
pulse is measured.
• When the measurement has been completed, the remai-
ning air pressure is released very quickly. The pulse, the
systolic and the diastolic blood pressure are displayed.
• The device requires approx. 5 seconds to store the deter-
mined measured value.
• You can interrupt measurement at any time by pressing
the ON/OFF button
.
• To switch off and release the pressure, press the ON/OFF
button
again. If you forget to switch off the device, it
switches off automatically after approx. 1 minute.
Wait at least 5 minutes before taking another measurement!
Evaluating results
Cardiac arrhythmia:
This instrument can identify possible cardiac arrhythmia
disorders during measurement and if necessary indicates
the measurement with the flashing icon
ecess
e
e
e
e
e
e
e
.
This may be an indicator for arrhythmia. Arrhythmia is a
condition where the heart rhythm is abnormal as a result
of defects in the bioelectrical system controlling the heart
beat. The symptoms (omitted or premature heart beats,
slow or excessively fast heart rate) may be caused, among
other things, by heart disease, age, physical predisposi-
tion, excessive use of stimulants, stress or lack of sleep.
Arrhythmia can only be ascertained through examination by
your doctor.
Repeat the measurement if the flashing icon
is displa-
yed after the measurement. Please note that you should
rest for 5 minutes between measurements and not talk or
move during the measurement. If the icon
an
n
n
n
n
n
n
n
d
appears
often, please contact your doctor. Any self-diagnosis and
Summary of Contents for BC 16
Page 52: ...52 Beurer 1 2 5 5...
Page 53: ...53...
Page 55: ...55 Pb Cd Hg 12 13 00 pm 1 00 5 1 5...
Page 56: ...56 50 2 40 Real Fuzzy Logic 5 1 5...
Page 57: ...57 5 National High Blood Pressure Education Pro gram Coordinating Committee...
Page 58: ...58 6 50 50 M 1 105 60 105 120 60 80 120 140 80 90 140 160 90 100 160 100 Adapted from JNC 2003...
Page 59: ...59 5 M 7 8 EE EE E1 20 E1 300 E2 E3 Er 9 BC 16 13 5 22 0 299 c 40 250 40 250 40 199 3...
Page 70: ...70...
Page 71: ...71...
Page 72: ...72 753 081 0608 Irrtum und nderungen vorbehalten...