![Beurer medical GL 48 Instructions For Use Manual Download Page 22](http://html1.mh-extra.com/html/beurer-medical/gl-48/gl-48_instructions-for-use-manual_2745022022.webp)
Beurer
GL48/GL49
23
If measurements are outside the specified range, check the following possible causes:
Cause
Action
• The first drop of control solution was not
disposed of.
• The tip of the bottle was not cleaned
correctly.
• The bottle was not shaken well enough.
Rectify the cause and repeat the test.
Control solution and/or test strips have
passed their expiry date or are contam-
inated.
Repeat the test using a new bottle of control solution
and/or new test strip from a new box.
The control solution, test strips or monitor
are too warm or too cold.
Bring the control solution, test strips and monitor to
room temperature (+20°C to +26°C) and repeat the test.
The check performed at room temperature is used as a
general functional check. The operating range specified
in the technical specifications is valid without restriction.
The test strips and control solution were
kept at a temperature and humidity out-
side the specified range.
Repeat the test using new correctly stored test strips
and control solution.
Damaged test strips.
For example:
• Test strips that were exposed to fresh
air for too long.
• The test strip box was not closed com-
pletely.
Repeat the test using a new test strip and/or correctly
stored test strips from a new box.
There is a problem with the monitor.
Contact Customer Services.
Functional test was incorrectly per-
formed.
Repeat the test and follow the instructions.
Warning
Do not use the system to measure your blood glucose level
if you are repeatedly provided with
measurements outside the specified range when using control solution. Contact Customer Services.
6. m
easurement
memory
For each measurement, your blood glucose value is automatically saved with the date and time unless
“CONTROL” was activated for a blood glucose measurement using control solution.
The memory can store a maximum of 900 measured values. If the memory is full, the oldest value is
replaced by the most recent value. You can call up every individual measured blood glucose value.
You can also calculate and display the average blood glucose value for the last 7, 14, 30 and 90 days.