GL48/GL49
13
3.3 Test strips
Front
Rear
3
2
1
1 Gap for blood input
2 Grip area
3 Contacts
You can identify the rear by
the contact tracks.
Insert the test strip into the de-
vice so that the contacts are
pointing inside the slot.
Make sure that the front of the
test strip is facing you.
Note
Read carefully the following information on handling and storing your test strips. The test strips will
only provide accurate measurements if all information is followed.
Warning
Use each test strip only
once
and for
one
patient only!
Handling test strips
Note
• Securely close the test strip box immediately after taking out a test strip.
• Do not use the test strips if they have expired . The use of expired test strips may lead to incorrect
measurements. The expiry date is located next to the hourglass symbol on the box.
• Use the test strip for measurement immediately after removing it from the box.
• You can touch any part of the test strip with clean, dry hands.
• Do not bend, cut or otherwise modify the test strips.
• Do not use test strips that have come into contact with fluids.