MasterCAL
Users Manual
66
7.
Press the
Ramp
softkey. The display changes to
the following:
gj41s.eps
8.
Fill in the parameters as prompted. Enter the
Start Value, End Value
, and
Ramp Time.
9.
To automatically stop ramping if a trip condition
is detected, set the
Trip Detect
to
Enabled
, and
select
Voltage
or
Continuity
as the trip function.
10.
Press the
Done
softkey. Note the
RAMP
annunciator next to
SOURCE
at the top of the
display.
11.
Select a low-to-high ramp or a high-to-low ramp
with the
Ramp Up/Down
softkey.
12.
To start ramping, press the
Start Ramp
softkey.
13.
Ramping continues until a trip is detected (if
enabled), the ramp time expires, or you press
the
Stop Ramp
softkey.