Specifications
107
DC Voltage Measurement
Range
Resolution
% of R% of Full Scale
1 Year
2 Year
110 mV
1
µ
V
0.025% + 0.015%
0.05% + 0.015%
1.1 V
10
µ
V
0.025% + 0.005%
0.05% + 0.005%
11 V
100
µ
V
0.025% + 0.005%
0.05% + 0.005%
110 V
1 mV
0.05% + 0.005%
0.1% + 0.005%
300 V
10 mV
0.05% + 0.005%
0.1% + 0.005%
Temperature Coefficient: (0.001% of rdg. + 0.0015% f.s.)/
°
C in the ranges -10 to 18
°
C and 28 to 50
°
C
Input Impedance: 5 M
Ω
Common Mode Error: 0.008% f.s./(Common Mode Volt)
Maximum Input Voltage: 300 V rms