Calibrator
Specifications
109
DC Current Measurement
Range
Resolution
% of R% of Full Scale
1 Year
2 Year
30 mA
1
µ
A
0.01% + 0.015%
0.02% + 0.015%
110 mA
10
µ
A
0.01% + 0.015%
0.02% + 0.015%
Temperature Coefficient: (0.001% of rdg. + 0.002% f.s.)/
°
C in the ranges -10 to 18
°
C and 28 to 50
°
C
Common Mode Error: 0.01% f.s./(Common Mode Volt)
Maximum Input Voltage: 30 V dc
Resistance Measurement
Range
Resolution
% of R ohms
1 Year
2 Year
11
Ω
0.001
Ω
0.05% + 0.05
0.075% + 0.05
110
Ω
0.01
Ω
0.05% + 0.05
0.075% + 0.05
1.1 k
Ω
0.1
Ω
0.05% + 0.5
0.075% + 0.5
11 k
Ω
1
Ω
0.1% + 10
0.1% + 10
Temperature Coefficient: (0.01% f.s. + 2 m
Ω)
/
°
C in the ranges -10 to 18
°
C and 28 to 50
°
C
Common Mode Error: 0.005% f.s/(Common Mode Volt)
Maximum Input Voltage: 30 V dc