In case there is a failure with the actuator or electronics and the user is left
suspended in mid-air, please follow the procedures below to safely lower the user
to a safer position.
Contact your dealer immediately if standard troubleshooting techniques do not
correct the failure. Do not attempt to lift until all failure and safety issues have
been resolved.
In case of lift failure, please follow the procedures below to safely lower the user.
The Emergency Lowering Device is located at the top of the actuator shaft. It is
intended for use if the actuator fail to operate while a patient is suspended.
The device consisting of a plastic collar ring that should be turned clock-wise
continually until the patient has been lowered.
(0(5*(1&</2:(5,1*0(&+$1,60
3/VHULHV
Turn clockwise to lower
24