or use cotton swabs with medical alcohol Wiping, rubbing with medical
alcohol can also have sterilization. Be careful not to add too much water or
alcohol to prevent damage to the inside of the product.
(2) Internal dirt: The lens of the internal probe is an important device. Do not
touch or press with your fingers or other objects, otherwise it will affect
the preparation of the measured result. When the lens surface is found to
be dirty, wipe the lens surface with a cotton swab moistened with 95% or
more absolute alcohol. Tip: Do not use 75% disinfectant alcohol to wipe the
lens (there will be traces of residual water). Do not use other chemical wipes
to wipe the lens (which will cause damage to the lens).
(3)Storage: Keep in a dry and dark place out of direct sunlight.
IX
Error message and solution
FAQ 1: LCD display message "HI"
Answer: Using infrared forehead temperature timer, the
LCD may display the message HI. The analysis shows that
the measurement range is exceeded: 1. The mode is
selected incorrectly; 2. The temperature exceeds 42.5
°C.
The following information appears when the message "L0" or "HI" appears:
Reason for the message "L0"
or "HI"
Suggestion
Temperature value is affected
by hair and sweat
Ensure that there are no obstructions
during temperature measurement
Temperature value is affected
by air flow
Ensure the air is stable during temperature
measurement
The measurement distance is
too far
Pay attention to measuring distance
(at 3 ~ 5CM)
Enter indoors from low or high
temperature outdoors
Please wait for 10 minutes and wait for the
subject's temperature to adapt to the
measurement environment before
measuring
FAQ 2: LCD display message "L0"
Answer: Using infrared forehead temperature timer, LCD may display
message L0.The analysis shows that it is lower than the measurement range:
1. The mode is selected incorrectly; 2. The temperature is lower than 32
°C
.