Chapter 7 Service
73
7.3
Check Measurement with Test Generator
You can use the test generator in code 60 to check the calibration
of the measurement. The test generator simulates the probe and is
enabled by entering a digit larger than 0.
It is advisable to simulate the empty count rate (code 20) as well
as the full count rate (code 21) with the test generator. Correct
switching of the alarm relay can be checked and the reading can
be presented on the display.
If the test generator is turned on:
the device acts as if the count rate were coming from the de-
tector.
the counts from the detector will be ignored.
the error relay is enabled to indicate that the readings do no
longer correspond to the level any more.
Note!
For the limit switch to continue working normally, you have to turn
off the test generator after the test.
To be on the safe side, the test generator will be turned off auto-
matically after 20 minutes.
The test generator is also disabled if the limit switch is locked by
entering the password.
Note!
We recommend documenting all settings in the start-up protocol
on the following pages.
7.4
Plateau Measurement
Plateau measurements are carried out to check the detector in
case of error (only for NaI detector with NaI crystal-multiplier
combination).
The function is started with code 68.
See also the section „Checking the Crystal-Multiplier Combination“
in the hardware manual.
Summary of Contents for LB4710-050
Page 2: ......
Page 4: ......
Page 8: ......
Page 32: ......
Page 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Page 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Page 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Page 44: ......
Page 58: ......
Page 98: ......
Page 104: ......
Page 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Page 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Page 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Page 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Page 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Page 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Page 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Page 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Page 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Page 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Page 120: ......
Page 141: ...Mini Switch LB 471 141 Notes...
Page 142: ...Mini Switch LB 471 142 Notes...
Page 145: ......
Page 147: ......
Page 149: ......
Page 151: ......
Page 161: ......
Page 195: ......
Page 215: ......
Page 219: ......
Page 223: ...Mini Switch LB 471 78 Notes...
Page 224: ...Mini Switch LB 471 79 Notes...
Page 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...