OPERON B 810 SURGICAL TABLE
GENERAL INFORMATION
Primary Hand Pendant Indicators
See FIGURE 2-6 for location of indicators
1.
AUX PENDANT
(orange) — Auxiliary hand
pendant drawer is open. The primary hand pendant
will still be enabled if the auxiliary hand pendant
drawer is open.
2.
SERVICE
(red) — Table malfunction. It is likely
that most of the functions still work properly and
can still be controlled from the primary hand
pendant. If the control pendant is not working, use
the auxiliary hand pendant and the manual foot
pump (if necessary) to articulate the table until the
malfunction can be repaired.
NOTE
NOTE
— If the SERV indicator lights,
call your local Berchtold Corporation
Service Representative.
3.
REVERSE
(orange)
—
Indicates reverse
orientation of controls.
4.
FLOOR UNLOCK
!
— Floor feet are raised so
table is mobile.
5.
FLOOR LOCK
"
— Floor feet are lowered so
table is locked in place.
6.
BATTERY
— Indicates charge status of the
batteries:
•
Green — Full to medium charge. Normal
Operation.
•
Yellow — Low charge. Table still operates
normally, but batteries should be charged at
the first opportunity.
•
Red (steady) — Recharge battery. Pump
motor is shut off and manual foot pump must
be used with auxiliary pendant.
•
Red (flashing) — Auxiliary pendant must be
used. Recharge battery immediately!
7.
ON
!
(backlight) — Control pendant is activated.
NOTE
NOTE
— In the event of a primary hand
pendant failure, use the auxiliary hand
pendant and foot pump to complete the
procedure and contact your local
Berchtold Corporation Service
Representative.
2
3
4
5
6
7
1
FIGURE 2-6
PRIMARY HAND PENDANT
INDICATORS
70-000-0026
Rev.
2