
OPERON B 810 SURGICAL TABLE
SPECIAL SAFETY INSTRUCTIONS
OPERATION
WARNING
WARNING
— The table should always
be locked to the floor whenever a
patient is on the table or being
transferred to or from the table. This is
accomplished by extending the table’s
feet by activating the FLOOR LOCK
function.
This table is not intended for
patient transport.
CAUTION
CAUTION
— Storage of clamps,
accessories, etc., on the base cover may
result in damage to lift column shields
and prevent proper elevation or descent
of the table.
CAUTION
CAUTION
— The footswitch’s
electrical connections are resistant to
fluid intrusion, but if fluid drainage is
anticipated, the footswitch should be
protected by wrapping it with a clear
disposable footswitch bag. This will
reduce the need to disinfect the
footswitch.
TABLE MOVEMENT WITH PATIENT
WARNING
WARNING
— If the table needs to be
moved with a patient on the table, it is
very important to follow the following
instructions to prevent patient injury or
equipment damage.
WARNING
WARNING
— The table should not be
used to transport a patient over a long
distance. This procedure is only meant
for repositioning the table within the
surgical suite.
WARNING
WARNING
— Do not mount knee
crutches or leg holders to the headrest
side rail. The headrest should only be
used for supporting the head or feet. Do
not lean on the headrest. The headrest
has a weight limit of 25lbs.
WARNING
WARNING
— Be sure that when
articulating the headrest, the locking
mechanism ends up properly engaged,
and not in between positions. The
headrest may move when load is applied
until the locking mechanism engages.
WARNING
WARNING
—
When releasing the
locking mechanism, support the
headrest, as it will fall freely if the
locking mechanism is not engaged.
WARNING
WARNING
— Failure to securely lock
the leg section into place may result in
patient injury. Always check the
retention of the leg section upon
installation.
ENVIRONMENTAL REQUIREMENTS
WARNING
WARNING
— The batteries are sealed
to prevent spilling of the acid
electrolyte. If punctured and contact is
made with the acid, flush with cold
water for 15 minutes.
WARNING
WARNING
— Although the batteries
are sealed to prevent spilling, they are
vented to allow for normal out gassing.
Because of this, the table should not be
used in a room that has no ventilation
system.
CAUTION
CAUTION
— The batteries contain
LEAD, and if replaced, need to be
recycled in accordance with local
environmental regulations.
CAUTION
CAUTION
— Do not expose the table to
steam or heat sources, or place the table
where it may be exposed to unusually
high temperatures.
CAUTION
CAUTION
— Allow the table to reach
room temperature prior to operation if
the table was exposed to excessively hot
or cold conditions during transport or
storage.
WARNING
WARNING
— When using the table in
the vicinity of high-frequency surgical
equipment, such as cardiac
defibrillators, and cardiac defibrillator
monitors, refer to the manufacturers’
instructions to ensure compatibility
issues are considered.
UNPACKING
WARNING
WARNING
— POSSIBILITY OF
INJURY: This table weighs
approximately 715 pounds (325 kg). At
least two persons are required to
remove the table from its shipping
container.
Rev. 2
70-000-0026
S-4