Troubleshooting

 

Symptom 

 

Possible cause and corrective action 

 

The system does not 
power on 

 

• 

Battery fully discharged. 

Connect the AC power adapter.   

• Power adapter does not function correctly. 

 

> Check if the power adapter has green light illuminated. 
> Check if the AC plug is connected firmly and correctly to the power 
adapter and matches the plug type of your country. 
> Check if the power adapter is connected firmly and correctly to the 
system. 

 

The system can’t charge 
or experience short 
runtime between charges 

 

Power supply is damaged/battery reaches end of life. 

 

> Allow the battery to charge overnight and check again the battery 
status. 

 

> Contact technical support.

 

Unsure of a function 
displayed in localized 
languages 

 

> Switch the system language back to English from 

Settings 

General 

Language

, and check again the function you 

want to use. 

 

> Contact technical support. 

 

The system can’t read/ 
write data from the 
microSD card 

 

• 

One or two USB storage devices are connected to the system 

at the same time.

> Remove all the connected USB storage devices from the system. 

 

• The microSD card is damaged. 

 

> Insert the microSD card into a computer for inspection. 

 

• The microSD card slot is damaged. 

 

> Insert another microSD card into the system for inspection. 

 

The system can’t read/ 
write data from the USB 
storage device 

 

• 

Two USB storage devices are connected to the system at the 

same time.

 

> Ensure the USB storage device you wish to read/write data from is 
connected to the system through the USB 3.0 port. 

 

• The USB storage device is damaged. 

 

> Insert the USB storage device into a computer for inspection. 

 

• The USB ports are damaged. 

 

> Insert another USB storage device into the system for inspection. 

 

No image or abnormal 
display on the system 
screen 

 

The system screen is not functioning. 

 

> Output the system display to an external monitor and check if 
images display normally on the external screen. 

 

> Contact technical support. 

 

Image Artifacts occur on 
the imaging screen 

 

• Electrical interference occurs. 

 

> Move the system away from any electromagnetic sources. 
> Remove the power adapter and other external devices (if any) 
from the system while keeping the transducer plugged for inspection. 

 

• The transducer connected is damaged. 

 

> Replace current transducer with another for inspection. 

 

Touch screen is 
unresponsive or 
misconfigured 

 

• Touch screen is damaged. 

 

> Inspect the panel surface carefully for cracks, cuts or any other 
damages. 

 

• Software malfunctions. 

 

> Connect a pointing device to the system through the USB port, and 
check if the pointer is displayed correctly on the screen. 

 

Software installation failed 

 

Check if the USB flash drives or the system’s USB ports are 
damaged. 

 

Summary of Contents for T3300v

Page 1: ...Vet Ultrasound System Quick Guide...

Page 2: ...Patient 3 Selecting a Preset 4 Splitting the Imaging Screen 5 Adding Annotations 6 Adding Body Mark 7 Adding Measurements 8 Saving and Printing the Image 9 Reviewing the Image V Using Image Controls 1...

Page 3: ...adily on green after the battery is charged l If the system is turned on the battery indicator lights off 5 Power Indicator Steadily on blue after the system enters Sleep mode 6 MicroSD card slot Inse...

Page 4: ...tilation slots Release excessive heat during operation to keep the system in a safe operating temperature 16 Power input socket Used to connect the system to power 17 Anti theft lock slot Used to lock...

Page 5: ...5 II Transducer...

Page 6: ...6 III Accessories Carrying the System Installing the Transducer Holder...

Page 7: ...7 Connecting the Transducer...

Page 8: ...8...

Page 9: ...9...

Page 10: ...he main screen or the user login screen after system startup Press and hold the Power button until the Power off menu appears on the screen then touch OK to turn off the system After turning on the sy...

Page 11: ...11...

Page 12: ...12...

Page 13: ...rol settings Scan mode buttons Open the next previous page of the image control settings Freeze button Freeze the current scan Save button Save a default set of image frames as a loop to the system ha...

Page 14: ...ing area displays as common usage Adjust the overall resolution Resolution High Resolution view a clearer yet superficial image General General Resolution view a general resolution image Penetration D...

Page 15: ...ge of the system storage used Storage usage empty Storage usage full Check and manage outgoing queues to the DICOM server DICOM Disconnected DICOM Connected Set current date and time Adjust the volume...

Page 16: ...o the next field To finish and close the keyboard touch 3 The system saves patient information automatically when you touch Start Exam Updating Patient Information 1 During real time scanning touch Pa...

Page 17: ...17 1 On the imaging screen touch Preset All the available presets compatible with the connected transducer display on the Preset screen...

Page 18: ...18 2 Touch the preset to scan and you will be redirected automatically to the real time imaging screen...

Page 19: ...19...

Page 20: ...l When in Review mode touch Show Annot to set whether to show all or the last annotations or to hide them all l To delete annotations including texts and arrows touch Erase Annot to erase the last an...

Page 21: ...xts directly then touch to close the keyboard 3 Touch Set home Label 1 Touch Label A predefined text menu appears at the control panel area 2 Select a text label to place it at the text home position...

Page 22: ...ing Settings BDMK Default BDMK for Application Edit select an option To delete the body mark added touch BDMK Erase Adding Measurements Measurements accompanying ultrasound images supplement other cli...

Page 23: ...option Measuring in B Color Power Modes Distance Measure a distance 1 Touch Distance A crosshair cursor appears on the image Touch anywhere on the imaging area to drag the target cursor to where you w...

Page 24: ...the second line extending out from the origin point and release it to place the end point of measurement The angle between the two lines is then formed Calculation Package Perform a set of predefined...

Page 25: ...rint out the current image Reviewing the Image On the real time or frozen imaging screen flick vertically on the thumbnail list to view the thumbnails of all the saved images loops To further examine...

Page 26: ...window and touch to select multiple images loops as expected To quickly select all images touch Select All To cancel all selections touch Deselect All 3 Touch an action button For example touch Delet...

Page 27: ...27 4 Using Image Controls B mode Image Controls...

Page 28: ...28...

Page 29: ...29...

Page 30: ...nd its amplitude or power The frequency shift is determined by the movement of the red blood cells relative to the transducer flow towards the transducer produces a higher frequency signal and flow aw...

Page 31: ...31...

Page 32: ...32...

Page 33: ...33...

Page 34: ...34...

Page 35: ...stem for inspection The system can t read write data from the USB storage device Two USB storage devices are connected to the system at the same time Ensure the USB storage device you wish to read wri...

Page 36: ...s are not correct Turn off the Bluetooth function then turn it back on The Bluetooth module is not functioning Connect another Bluetooth device to the system for inspection DICOM connection failed The...

Reviews: