07/ 2016
BENNING MM 7-1
20
7.9 Capacity ranges
Conditions: capacitors discharged and connected in accordance with the polar-
ity stated.
Overload protection: 1000
V
AC/ DC
Measuring range
Resolution
Measurement accuracy
1 µF
1 nF
± (1.2 % of r 5 digits)
10 µF
10 nF
± (1.2 % of r 5 digits)
100 µF
100 nF
± (1.2 % of r 5 digits)
1 mF
1 µF
± (1.2 % of r 5 digits)
10 mF
10 µF
± (1.2 % of r 5 digits)
Max. measuring time:
0.7 seconds for 1 nF - 1 mF
3 seconds for 1 mF - 10 mF
7.10 Frequency ranges
Overload protection: 1000
V
AC/ DC
Measuring range
Resolution
Measurement accuracy
100 Hz
0.01 Hz
± (0.1 % of r 2 digits)
1 kHz
0.1 Hz
± (0.1 % of r 2 digits)
10 kHz
1 Hz
± (0.1 % of r 2 digits)
100 kHz
10 Hz
± (0.1 % of r 2 digits)
Minimum frequency: 1 Hz
Minimum sensitivity: > 5 V
SS
for V
AC
1 Hz - 10 kHz
>
10
V
SS
for V
AC
10 kHz - 100 kHz
>
2
mA
SS
for mA
AC
>
0.2
A
SS
for A
AC
7.11 Temperature ranges °C/ °F
Overload protection: 1000
V
AC/ DC
Measuring range
Resolution
Measurement accuracy*
- 40 °C to 400 °C
0.1 °C
± (1 % of r 30 digits)
- 40 °F to 752 °F
0.1 °F
± (1 % of r 54 digits)
* The measuring accuracy of the K-type temperature sensor has to be added
to the specified measuring accuracy.
Wire temperature sensor (type K):
Measuring range: - 60 °C to 200 °C
Measurement accuracy: ± 2 °C
The measuring accuracy applies to stable ambient temperatures < ± 1 °C. After
a change of the ambient temperature of ± 2 °C, the measuring accuracy data
will apply after 1 hour.
7.12 PEAK HOLD
DC/ AC V Measuring range
Measurement accuracy
60 mV
± (0.08 % of r 155 Digit)
600 mV
± (0.08 % of r 152 Digit)
6 V
± (0.08 % of r 152 Digit)
60 V
± (0.08 % of r 152 Digit)
600 V
± (0.08 % of r 152 Digit)
1000 V
± (0.08 % of r 152 Digit)
DC/ AC A Measuring range
Measurement accuracy
60 mA
± (1.2 % of r 153 Digit)
600 mA
± (1.2 % of r 153 Digit)
6 A
± (1.2 % of r 153 Digit)
10 A
± (1.2 % of r 153 Digit)
8. Measuring with the BENNING MM 7-1
8.1 Preparation for measuring
Store and use the BENNING MM 7-1 only under the correct temperature condi-
tions stated. Always avoid longer exposure to sunlight.
- Check the rated voltage and rated current stated on the safety measuring
leads. The safety measuring leads supplied with the unit are suitable for the