Chapter 5
Control Solution Test
Control solution is used to check the performance of the kit. The meter performs adequately
when control test falls within the range listed on test strip vial label.
The meter kit should be check:
• When the meter and strip do not work properly.
• When the test result is unusual or inconsistent.
Control solution range is shown as follow:
Please check your strip vial label for exact range.
Directions:
• Insert an unused strip
into the meter.
1
2
• Shake well and discard
the first three drops of
control solution. And
put one drop onto a
clean surface.
3
4
• Touch the control solution with
strip sample inlet .
• The result will display
after countdown.
• Compare the result with
the range listed on test
strip
vial.
5
Note:
• Do not reuse the test strip.
• Newly opened vial of control solution and test
strips must be marked with the opening date.
• Close the strip vial lid completely after use.
• Do not use the meter if the control range do
not fall in the correct range.
• If the control test keeps result wrongly,
please contact an authorized distributor for
help and service.
• Control solutions are not included. Please
contact an authorized distributor for checking
the meter kit performance.
19
20
• Press “M” button for 3 seconds
to enter the control solution
mode.
Test Strip
Operation
Storage &
Transportation
Storage &
Transportation
90%
10%
98-164 mg/dL
5.4-9.1 mmol/L
EA13120901
2017-01
98-164 mg/dL
5.4-9.1 mmol/L
G1303190
2017/01
EA13120901
2017-01
control range level 0/ level 1/ level 2
Operation
Storage &
Transportation
10
Storage &
Transportation
90%
10%
New Taipei City 242
MDSS GmbH
Schiffgraben 41
30175 Hannover, G
FIG-80-SLG10-6BK
Control Range
Operation
Storage &
Transportation
Storage &
Transportation
90%
10%
98- 164 mg/dL
5.4-9.1 mmol/L
level 0 / level 1 / level