Function
19
isoHV525_D00297_04_M_XXEN/05.2018
3.2.2.7
Interface/protocols
The ISOMETER® uses the serial hardware interface RS-485 with the following protocols:
BMS
The BMS protocol is an essential component of the Bender measuring device interface
(BMS bus protocol). ASCII characters are used in the data transfer.
Modbus RTU
Modbus RTU is an application layer messaging protocol and it provides Master/Slave
communication between devices that are connected altogether via bus systems and
networks. Modbus RTU messages have a 16-bit CRC (Cyclic Redundant Checksum),
which guarantees reliability.
IsoData
The ISOMETER® continuously sends an ASCII data string with a cycle of approximately
1 s. Communication with the ISOMETER® within this mode is not possible and no addi-
tional transmitter may be connected to the RS-485 bus cable. The ASCII data string for
the ISOMETER® is described on
.
The isoHV525-S4-4 ISOMETER® uses the device name "isoHV425" when it
reports on the BMS bus.
The isoHV525-S4-4 ISOMETER® uses the device name "isoHV425-D4-4"
when it reports on Modbus.
The IsoData protocol can only be terminated by sending the command
"Adr3" during an interval of the isoHV525-S4-4 ISOMETER®.