![Bender ISOMETER isoBAT425 Manual Download Page 11](http://html.mh-extra.com/html/bender/isometer-isobat425/isometer-isobat425_manual_2731653011.webp)
Function
Function
isoBAT425_D00308_04_M_XXEN/07.2020
11
3.2.6.3
Battery voltage
The nominal system voltage
U
n
is monitored with the voltage comparators described
above. The insulation fault
R
C
between the unconnected strings as well as the completion
of the battery can be detected via these comparators.
3.2.7 Error codes
If, contrary to expectations, a device error should occur, error codes will appear on the
display. Some of these are described below:
Internal device errors E.xx can be caused by external disturbances or internal hardware
errors. If the error message occurs again after restarting the device or after a reset to
factory settings (menu item "FAC") , the device must be repaired.
After eliminating the fault, the alarm relays switch back
automatically or by pressing the
reset button.
Error code
Meaning
E.01
PE connection error
The connections "E" or "KE" to earth are interrupted.
Action:
Check connection, eliminate error. The error code will be erased
automatically once the error has been eliminated.
E.02
System connection fault
The internal resistance of the system is too high, the connection "L+"
or "L-" to the system is interrupted or the DC system to be monitored
has the wrong polarity at
U
n
< -10 V.
Action
:
Check connection, eliminate error. The error code will be erased
automatically once the error has been eliminated.
E.05
Measurement technique error
Calibration invalid for current software version
E.07
The permissible system leakage capacitance has been exceeded.
System leakage capacitance
C
e
> 4 μF
E.08
Calibration error during the device test
Action:
If the error continues to exist after checking the device connections,
there is an error inside the device.
The self test can take a few minutes. It can be suppressed for the duration of the device
start by setting the parameter "S.Ct = off" in the
. This allows the ISOMETER® to
enter measurement mode quickly after connecting the supply voltage
U
s
.
3.2.8 Connection monitoring
3.2.8.1
E-KE monitoring
The E-KE monitoring checks the connection between terminals
"
E
"
and
"
KE
"
and thereby
ensures the connection to earth of the ISOMETER® and the battery enclosure. An error is
reported with the code
E.01
and device error. It operates constantly in the background
and does not affect the measuring function of the device.
3.2.8.2
System connection monitoring
The system connection monitoring monitors the connection of terminals
"
L+
"
and
"
L-
"
to
the battery by means of the coupling impedance. An error is reported with the code
E.02
and device error. The monitoring requires a resistance smaller than 12 kΩ and a system
leakage capacitance
C
e
larger than 80 nF between terminals
"
L+
"
and
"
L-
"
. In the case of
separated battery strings, the coupling impedance ensures that these minimum requi-
rements are fulfilled. In the case of closed battery strings, the battery provides the requi-
red connection between
"
L+
"
and
"
L-
"
with its low-ohmic internal resistance. The system
connection monitoring operates constantly in the background and does not influence
the measuring function of the ISOMETER®.
Above a measuring pulse period duration of 2.0 s, a missing connection to "L+" or "L-" is
only reliably detected under the conditions specified in
.
The system connection monitoring can be deactivated by means of the "nEt" parameter
in the
.
3.2.9 Reset and test
3.2.9.1
External test/reset input and stop mode
The fault memory (
"
M
"
) is deleted by briefly connecting (0.2 s < t < 1.5 s) the terminal
"
T/R
"
to earth (terminal
"
E
"
). If the connection time exceeds 1.5 s, the manual self test
starts. If the terminal
"
T/R
"
is still connected to earth after the device self test has been
completed, the ISOMETER® changes to stop mode ("StP" indication on the display) and
the measuring function is deactivated.