H5000 Series MultiTherm Shaker
s
—
15
—
d)
Repeat the above step for the remaining
temperature calibation points.
Following the adjustment to the final calibration point,
the temperature calibration of the instrument has now
been completed and the displayed temperature will
accurately represent the temperature of the heating
block.
NOTE: During the temperature calibration, press the “
” and “
” Prog. keys
simultaneously to cancel the calibration. The system will default back to the former
calibration.
P
:
40.1 ADJ
AdjTemp= 40.0
P
:
100.0 Adj
AdjTemp= 98.0