H5000 Series MultiTherm Shaker
s
—
14
—
Fig. A
Notes: Please read the actual value after allowing 20 minutes constant temperature
equilibration time.
To enter the temperature calibration mode, press
the “
” and “
” Prog. Keys simultaneously.
The instrument begins to cool to the low
calibration point.
- H5000-H
Calibration points: 40°C, 100°C
- H5000-HC
Calibration points: 10°C, 40°C, 100°C
c)
After 20 mintues
,
read the temperature of the
Thermometer (for ex. 9.8°C). Using the “
”
or “
” Temp. keys, adjust the temperature of
the display to match that of the thermometer
and press the “start/stop”key to confirm.
Thermometer
Insulating sheet
Standard Mineral Oil
Well-bottom
P
:
20.5 ADJ
AdjTemp= 10.0
P:10.0 ADJ
AdjTemp= 9.8
Press “stop/start”
P:10.0 ADJ
AdjTemp= 10.0