H5000 Series MultiTherm Shaker

s

 

 13 

 

b)  To deactivate the shaker speed, press and hold the “

” Speed key until the set 

speed shows “OFF”.    The shaker is now deactivated and the instrument can be 
used solely for temperature control. 

c)  To deactivate the timer, press and hold the “

” Time key until the set time shows 

“OFF”.  The timer is now deactivated and the instrument will run continuously 
until the user presses and holds the start/stop key. 

NOTE:    When running consecutive programs in the Seq. Link mode, the time should 
not be set to “OFF” in any program. 
 

4.

 Pulse

 

By pressing and holding the Pulse key, the 

instrument mixes at the selected speed for as 

long as the Pulse key is held down. The time is 

counted in seconds until 999 seconds has 

expired. 

 

5.

 Temperature calibration

 

 

The temperature of the Instrument has been factory adjusted and calibrated prior to 

being shipped. But if there is deviation between the actual temperature and the displayed 
temperature, the user can self-calibrate the instrument in order to correct the error.   

 

Note: The temperature accuracy will be within ±0.5°C.   
 

 
Calibration Procedure:       
a)  Verify that the ambient temperature is less than 25°C and power up the instrument.  

Once powered on, verify that the current temperature in display is below 25°C. If the 
temperature is above than 25°C, please wait for the unit to cool down to 25°C or less.   

b)    Inject olefin oil into one of the cone-shaped wells, and insert a thermometer into this 

well. (Make sure the accuracy of the thermometer is to be within 0.1°C and that it is 
completely immersed into well). Heat insulation material is needed on the block to 
insulate it from the ambient temperature.    See Fig a.: 

 
 
 
 
 
 

Pulse   is   running

 

   

600rpm     023S 

Summary of Contents for H5000-H

Page 1: ...MultiTherm Shaker Operations Manual Item H5000 H and H5000 HC...

Page 2: ...instrument properly please read this manual carefully before operation Please keep it available for future use Prior to the initial operation Please check the Instrument and the accessories against th...

Page 3: ...elines Use this product indoors Do not operate in a damp or wet environment Do not attempt to disassemble or repair the Instrument by yourself Disassembly will void the warranty If there is a malfunct...

Page 4: ...environments Keep the instrument and the air vents at least 4 inches from walls other instruments or any obstructions Shut the power off when you finish your work Unplug the machine if it will not be...

Page 5: ...8 1 Physical Description 8 2 Keyboard and Display panel 9 3 Key Functions 10 Chapter 4 Operation Guide 11 1 Single Temperature Speed and Time setting 11 2 Sequence Link Setting 12 3 Disabling the Temp...

Page 6: ...oresis serum solidification etc Features of this product as follows 1 Because it can be used with various optional mixing blocks the instrument can adapt to different tubes and plates to accommodate m...

Page 7: ...C Cooling time N A Block Configuration A BLOCK 96 well Plate or 96 x 0 2ml B BLOCK 0 5ml x 54 C BLOCK 1 5ml x 35 D BLOCK 2 0ml x 35 E BLOCK 15ml x 12 F BLOCK 12mm x 24 G BLOCK 50ml x 6 H BLOCK 0 5ml x...

Page 8: ...to 100 C Cooling time 30min from room temp to 20 C below room temp Block Configuration A BLOCK 96 well Plate or 96 x 0 2ml B BLOCK 0 5ml x 54 C BLOCK 1 5ml x 35 D BLOCK 2 0ml x 35 E BLOCK 15ml x 12 F...

Page 9: ...cal description location of controls function of individual keys as well as some preparations before the initial power on You should be familiar with this chapter before the Thermo Shaker is first ope...

Page 10: ...Shakers 9 2 Keyboard and Display panel Display Panel P1 100 0 1200 10 30 P1 100 0 1200 15 00 Current temperature Current speed Remaining time Selected temperature Selected speed Selected time Current...

Page 11: ...and hold the or keys Sequence Link Key Press the Seq Link key to run consecutive programs Choose from P1 to P2 P1 to P3 P1 to P4 or P1 to P5 to run up to five consecutive programs The Sequence link fu...

Page 12: ...rature The initial preset mixing speed and time are also displayed at 1000rpm and 10 00 respectively c Five programs are available for storage By pressing the or Prog keys choose the desired program P...

Page 13: ...To adjust the ending program press the or Prog Keys Once the ending program has been selected press the Seq Link key to finalize c To begin the first program press the start stop key The first progra...

Page 14: ...ture calibration The temperature of the Instrument has been factory adjusted and calibrated prior to being shipped But if there is deviation between the actual temperature and the displayed temperatur...

Page 15: ...low calibration point H5000 H Calibration points 40 C 100 C H5000 HC Calibration points 10 C 40 C 100 C c After 20 mintues read the temperature of the Thermometer for ex 9 8 C Using the or Temp keys a...

Page 16: ...temperature calibration of the instrument has now been completed and the displayed temperature will accurately represent the temperature of the heating block NOTE During the temperature calibration pr...

Page 17: ...rench 3mm to remove the four screws that hold the block to the heating board b Close the lid and lift the block up and away from the base unit c Place the desired block softly on the heating platform...

Page 18: ...atures are significantly different Temperature sensor failure Call for service 3 OPEN in the temperature display with a short beep Temperature sensor failure or the environmental temperature is below...

Page 19: ...Notes PH 908 769 5555 EM info benchmarkscientific com...

Reviews: