3. Introduction
3-1 Precautions for Installation
Keep the equipment away from water.
•
Keep in a circumstances safe from infl uance by temparature, humidity, wind, sun light, air containig salts and
•
minerals.
Care about stability such as inclination, vibration and inpact, including handling and transportation.
•
Do not keep the equipment in a place where chemicals are or where gas is emitted.
•
The fl oor construction required to safely support the chair and delivery system is 105 lbs./ft2 (500kg/m2 ) at
•
minimum.
During lifting and unpacking of the chair, make sure to hold only the designated parts.
•
Do not drop or hit the chair.
•
After the chair is unloaded from the palette and placed at the desired location, please make sure to remove the
•
shipping bolt with the red tag. Damage to chair may occur if shipping bolt is not removed prior to chair opera-
tion.
Do not connect to power supply other than AC120V 60HZ.
•
Ground chair properly prior to turning power on.
•
Chair base must be anchored to the fl oor for maximum stability.
•
Refer to the installation manual of the chair and dental unit (if it is to be used) prior to, and during installation.
•
When the installation process has been completed, verify that all the mechanical and electrical functions are
•
working properly and that there is no evidence of oil leakage.
Protective footwear and thick gloves are highly recommended at unpacking.
•
Do not modify this equipment.
•
In the case separation from any electrical poles is necessary, pull of the power plug from the power supply
•
outlet.
3-2 Required Tools
The following tools are necessary for installation of the chair.
Phillips screw driver #2
Allen key wrenches, metric size
1) Place chair carton close to installation location. Remove all the
staples fi xing the carton to pallet, or cut the carton just above the
stapled line, and remove the carton. (See Fig 4-1-1)
2) Using a 10mm socket, loosen and remove two packing screws
holding the chair base into pallet. (See Fig 4-1-2)
Staples
Packing Screws
Fig 4-1-1
Fig 4-1-2
4. Installation
4-1-1 Unpacking the chair base
Do not connect in the power supply line before removing the red
tagged shipping bolt.
CAUTION
3