background image

1. OVERVIEW, MAJOR COMPONENTS AND SYMBOLS

1-1. Overview and major components

10

17

18

13

14

15

16

11

12

2

1

3

4

5

6

7

8

9

1. Headrest Assembly
2. Backrest Cushion
3. Sling
4. Armrest  
5. Side Frame
6. Rotation lock lever
7. Main Link Cover

Seat Raise

Seat Lower

Backrest Raise

Backrest Recline

Backrest Raise

Auto Return

Pre-set

Backrest Recline

Seat Raise

Seat Lower

Seat height and Backrest 
Control on backrest

Foot Control

Touchpad (Optional)

A

P

Backrest
Recline

Seat Raise

Backrest Raise

Seat Lower

Auto Return

Pre-set

8. Safety Plate
9. Base Cover
10. Seat Cushion
11. Pump Cover
12. Power Supply Cable
13. Backrest Cover
14. Rear Flange Cover

15. Seat height control
     (Seat Raise/lower)
16. Backrest control
     (Backrest Raise/recline)
17. Fuse holder
18. Main Power Switch
 

1

1-2. Symbols

Applied part complying with the specifi ed requirements of IEC60601-1:2005 to provide protection against elec-
trical shock, particularly regarding allowable patient leakage current and patient auxiliary current.

Caution, warning or note

Summary of Contents for BEL-20+

Page 1: ...l provides installation instructions for the BEL 20 Chair The instructions contained in this booklet should be thoroughly read and understood before installation of the chair After the installation ha...

Page 2: ...psi 600hpa 1060hpa Classification a Protection against electric shock Class I Equipment Type B Applied Parts Headrest Backrese and Seat cushions b Protection against water ingress Chair IPX0 Foot cont...

Page 3: ...est Foot Control Touchpad Optional A P Backrest Recline Seat Raise Backrest Raise Seat Lower Auto Return Pre set 8 Safety Plate 9 Base Cover 10 Seat Cushion 11 Pump Cover 12 Power Supply Cable 13 Back...

Page 4: ...e for relays 0 5A 125V Current rating 35A at 250VAC Fast blow Mode of Control Foot switch and backrest control Maximum Patient Load 300 lbs 135 kg Weight 333 lbs 150 kg 2 2 Specifications 8 20 10 10 2...

Page 5: ...chored to the floor for maximum stability Refer to the installation manual of the chair and dental unit if it is to be used prior to and during installation When the installation process has been comp...

Page 6: ...le tie 4 x Pan head screw M6 x 15mm 2 x Round head screw M5 x 10mm 2 x Round head wood screw M4 x 10mm 2 x Round head screw M3 x 10mm 5 x Flat head screw M5 x 10mm 5 x Plastic washers and caps 2 x Rou...

Page 7: ...arton box containing Armrests headrest mechanism and bag of hardware To avoid the risk of electrick shock this equipment must only be connected to a supply mains with protective earth Grounding reliab...

Page 8: ...ch headrest cushion to headrest mechanism and slide headrest bar into opening in top of backrest See Fig 4 2 5 10 Locate the red warning tag which is attached by a string to the rubber air vent plug i...

Page 9: ...krest adjustment screw clockwise See Fig 4 3 1 Solenoid Valve Block Seat Lower Screw Backrest Recline Screw Adjustment of the other screws for seat raising and backrest raise is not recommended NOTE 1...

Page 10: ...Seat Limit 1 Slide the Normal Limit switch to the Limit position 2 Move the seat base to the desired lowest position See Important box on above 3 Press and release the Store button on the PCB 4 Press...

Page 11: ...en Blue Black White Pink Gray Pink Gray Pink Gray Black COM Green Back Forward Down Up Brown Yellow Orange Red Red Brown Yellow Orange Green Down Up Back Up Down COM Forward Back Forward COM Red Yello...

Page 12: ...S A TEL 732 469 5000 800 223 1192 Fax 732 526 6322 800 280 7504 TAKARA CO CANADA LTD 2076 S Sheridan Way Mississauga Ont L5J2M4 Can TEL 905 822 2755 Fax 905 822 6203 TAKARA BELMONT CORPORATION Manufa...

Reviews: