SOLARIS  Biological  Microscope  operation  manual. 

 

7

Fig.10 

2

1

6 . Operate coarse and fine focusing control knob   

The coaxial coarse and fine focusing system is adopted in this instrument, with coarse tension adjusting device. 

(1)    The control knob 1 is for fine focusing, the control knob 2 is for coarse focusing

Turn the coarse focusing control 

knob 2 anticlockwise to lift the objective and clockwise to lower it. As shown Fig.10 

(2)    Put the 10X objective into the light path with the turn of nosepiece (When turn into the right position, the objective 

will has a automatic halt.). 

(3)    Turn coarse control knob unit the stage in the highest position, observe through the eyepieces, slowly turn coarse 

control knob, with the stage slowly down, to focus until the image clear and stop turning. 

(4)    Turn fine control knob to precision focus for a clearly image. 

(5)    The  knob  3  is  to  lock  the  stage  position  of  height;  this  function  is  making  sure  the  long  working  distance  of 

objective  not  to  be  touched  with  stage  or  simple.  When  the  stage  be  locked  in  the  preset  position  with  turn  the 

coarse, you can rapidly focus the image clear.       

 

 

 

 

 

 

 

 

 

 

 

 

 

(6)      The tension of the coarse focus control knob 2 is adjustable and preset at the factory for ease of use. If you wish to 

adjust the coarse focus tension, turn the tension knob 4 to adjustment. Turn the wheel anticlockwise increase the 

tension, and clockwise decreases it, as indicated by the arrow in the figure 8. 

 

 

 

Too high tension may be affected operation and physical discomfort. 

 

Caution

 

When  you  use  the  high  magnification  objective,  please  use  the  10X  objective  focus  the  image 

clear at first and set the knob 4 , then turn to observe by the high magnification objective, you can 

make  the  stage  to  the  preset  position  by  the  coarse  control  knob,  and  make  the  more  precision 

focus by the fine control knob 

Reminder

 

Summary of Contents for SOLARIS

Page 1: ...SOLARIS Biological Microscope Operating Manual ...

Page 2: ...ife of this unit and maintain it properly please read this manual carefully before operating this unit Caution This manual uses the following symbols for safety reminders Be sure to observe these warnings in order to operate this unit properly and safely Caution Negligence in heeding the caution of this symbol may affect the viewing performance of this unit Provide instructions and skills in opera...

Page 3: ...or dampness environment for a prolonged period of time A suitable operating environment is designated at a temperature of 5 C 35 C and relative humidity of 20 80 at 25 C If water splashes on this unit turn off the power switch and remove the power cord immediately and then wipe the water off with dry cloth When any foreign object enters or drips onto this unit please stop operating the unit and co...

Page 4: ...rial Wipe off any dirt on the surface of the bulb with a clean cotton fabric dipped in alcohol If the dirt is not thoroughly removed it would etch the surface of the bulb weakening its brightness and shortening its life Mount the bulb with care to avoid slipping off or injuries to your fingers When replacing the bulb make sure its contact is intact If its contact is damaged the bulb may be disable...

Page 5: ...e of Contents I Characteristics and applications of this unit 1 II Structural features of this unit 1 III Installation of this unit 2 IV Technical specifications 3 V Operation 4 VII Maintenance 10 VIII Troubleshooting Errore Il segnalibro non è definito ...

Page 6: ...it SOLARIS biological microscope is equipped with eminent UIS optical system provides excellent optical performance and the update of the system Because of the streamline figure and pastel color it has more beautiful appearance It achieves the notion of ergonomics design so it can provides the more comfortable operation and operational room It can be used in the field of the biology medicine indus...

Page 7: ...inocular locates the frontispiece of main body 4 Install two eyepieces 4 into eyepiece tube turn the eyepieces for about a circumference to make the eyepieces fit the eyepiece tube 5 Take out the condenser according to Abbe and turn down the support of condenser put the condenser according to Abbe into the support of condenser 6 Connect power supply to the power outlet of main body 7 Check the rel...

Page 8: ...ical Stage size 193 X 155mm Moving Range 75 X 50 mm Light Source Halogen Bulb 6V 20W adjustable of the brightness Power Supply AC Voltage 85V 265V 50 60Hz fuse parameter 250V 3 0A Eyepieces Centrimg Telescope Infinity phase contrast plan achromatic 10X0 25 PHP Infinity phase contrast plan achromatic 20X0 40 PHP Infinity phase contrast plan achromatic 40X0 65 PHP spring Objective Infinity phase con...

Page 9: ...agm6 little compare with field as shown in Fig 4 a if the field diaphragm not in the center of field it can be adjusted by the two inner hexangular wrench s to the center of field as be shown in Fig 4 b Restart the field diaphragm to larger than field shown in Fig 4 c Caution Don t keep the brightness control knob at the brightest position for a prolonged period of time otherwise the life of the b...

Page 10: ...d and turn the Condenser Adjustment to the position of under the simple or little more high of it do it all of above and lock the screw as be shown in the Fig 6 About the adjustment and observation of the phase contract condenser operation please read the The Manual of the Phase Contract Condenser Operation Check the position state of observation photography switch pole 1 Push the pole into observ...

Page 11: ... of microscopes have N 5 diopter for you adjustment 4 If you choice the right eyepieces as the calibration reference please reset the diopter as above 1 3 steps just change the relative eye to another eye 5 Adjustment of interpupillary distance Parallax can be eliminated by adjusting the interpupillary distance so that the distance of the eyepiece tube is identical with interpupillary distance and...

Page 12: ...ision focus for a clearly image 5 The knob 3 is to lock the stage position of height this function is making sure the long working distance of objective not to be touched with stage or simple When the stage be locked in the preset position with turn the coarse you can rapidly focus the image clear 6 The tension of the coarse focus control knob 2 is adjustable and preset at the factory for ease of ...

Page 13: ...en the center and direction of the image a synchronization adjustment will be necessary as follows a Push in push pull rod 4 observe with eyepieces Find a feature point in the field of view a readily identifiable target such as S point in Fig 12 a move it to the center of the field of view If there is a division eyepiece move the target to the reticle intersection of the division eyepiece as shown...

Page 14: ...e screws4 and pull out the lamp house cover take out the bad bulb3 and place the new one 4 If you want to adjust the position of the bulb you can Loosen the adjustment screw and move the bulb Front and back as far as the appropriate position you want finally close the screw NOTE Above steps not for the LED illumination apparatus 2 Replace fuse The fuse is installed in the fuse socket 1 as shown in...

Page 15: ... gently the mix ratio is alcohol 20 30 and ethyl ether 70 80 4 Cleaning the surface of this unit Wipe it with clean soft cloth heavy dirt may be wiped off with a neutral detergent 5 Keeping When this unit is not to be used for a long time turn off the power supply of this unit allow the bulb to cool down sufficiently put on the dust shield store this unit at a dry ventilated and clean place free f...

Reviews: